Immunity test
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
RATED power frequency
magnetic fields IEC
61000-4-8
Electrical fast transient/
burst
IEC 61000-4-4
Surges
IEC 61000-4-5
Conducted disturbances
induced by RF fields
IEC 61000-4-6
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Note: UT is the AC mains voltage prior to application of the test level. If in the unlikely event the Philips
Avent single/double electric breast pump Advanced is locked inoperative in a function by
disturbances from other devices beyond the levels stated in the table above, the device will shut-off
completely and restart. In that case make sure the distance to the other device causing this
disturbance is enlarged.
Immunity test level
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
30 A/m
50 Hz or 60 Hz
±2 kV
100 kHz repetition
frequency
±1 kV Line-to-line
±2 kV Line-to-ground
3 Vrms, before
modulation is applied
0.15 MHz – 80 MHz
6 Vrms, before
modulation is applied, in
ISM and amateur radio
bands between 0.15 MHz
and 80 MHz
80 % AM at 1 kHz
0 % UT; 0.5 cycle at 0°,
45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle and 70 %
UT; 25/30 cycles single
phase at 0°
0 % UT; 250/300 cycle
Electromagnetic environment -
guidance
The device is immune against any RF wireless
communication device such as smart phone,
Wi-Fi or Bluetooth device at a distance of
30 cm (11.8 in) or more. Avoid a closer
distance as this may result in improper
operation.
Mains power quality expected of a typical
commercial or residential environment.
Mains power quality expected of a typical
commercial or residential environment.
Mains power quality expected of a typical
commercial or residential environment. At
power interruptions the breast pump will
switch off.