INSTRUCTIONS FOR USE
urology, laparoscopy, general and robotic surgery. These devices are intended to
be used by healthcare professionals within the Operating Room setting.
2.6 Residual risk:
This product complies with relevant performance, safety standards. However,
patient harm from misuse, device, function or mechanical hazards cannot be
completely excluded.
3. Equipment Setup and Use:
3.1 Prior to use:
a. Inspect the product looking for any visible damage or sharp edges that could
be caused by a drop or impact during storage.
b. Make sure product has been properly cleaned and disinfected and wiped dry
prior to each use.
3.2 Setup:
Prior to setting up the Stirrups, assess the patient for any pre-existing conditions that
might contraindicate the use of leg positioning devices.
1. Position the Rail Clamps adjacent to the patient's hip.
2. Attach the Stirrup to the table rail by sliding the Stirrup Blade Mount into the
slot on the top of the Rail Clamp. Left and Right Stirrups must be in the exact
same location on opposite sides of the OR table.
NOTE: Ensure Left and Right Stirrups are on the proper side; the Boots are
marked (on the bottom of the Boot) for the patient's left and right.