8. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries. The cuff line
should not be bent sharply.

9. Rectifying faults

In the event of errors, the
display. Error messages may appear if:
• it was not possible to record the pulse: E1;
• you move or speak during the measurement: E2;
• the cuff is fastened too tightly or loosely: E3;
• errors occur during the measurement: E4;
• the pump pressure is higher than 300 mmHg: E5;
are almost empty: E6.
• the batteries
In such cases, repeat the measurement. Ensure that the cuff
line is correctly attached and that you do not move or speak. If
necessary, reinsert or replace the batteries.

10. Specifications

Model no.
BM 45
Type
M1501
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
error message appears on the
Measurement range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
and transport condi-
tions
Power supply
Battery life
Classification
28
Cuff pressure 0 – 300 mmHg,
systolic 60– 250 mmHg,
diastolic 40 –199 mmHg,
Pulse 30 –180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 160 mm x W 107 mm x H 50 mm
Approx. 408 g
(without batteries, with cuff)
22 to 36 cm
+5 °C to + 40 °C, 15 – 90 % relative air
humidity (non-condensing)
- 25 °C to + 70 °C, ≤93 % relative air
humidity, 700hPa – 1060hPa ambient
pressure
4 x 1,5 V
AA batteries
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF (cuff), CLASS II (when using a
mains part)