– It is not permitted to change the
adjustment of 1 through 5 while a
patient is sitting in the CPM device.
– Movements must not cause any
pain or irritation.
– Patients must be fully conscious
while being instructed in the use of
the CPM device and during therapy.
– The choice of the therapy parame-
ters to program and of the therapy
protocols to use is restricted to
the responsible physician or
therapist.
It is the physician's or therapist's
decision whether or not to use the
CPM device on a specific patient
– The patient must be familiar with the
functions of the OptiFlex
gramming unit and the unit must
be within easy reach of the patient,
allowing him or her to stop therapy,
if needed. Patients unable to
operate the programming unit ,
e. .g. paralytic patients, must never
be left unattended during therapy.
– Write the patient's name on the
patient chip card. The card should
only be used for this patient. If the
patient chip card is used for another
patient, be sure to delete the
previous patient's data from the
card first (see sections 4.1 and 5.3,
paragraph "New Patient").
Use original chip cards only.
– All accessories used with the
OptiFlex
®
S device must first be
approved by CHATTANOOGA.
– Do not allow parts of the body or
any objects (such as blankets,
cushions or cables) to get caught in
the moving parts of the CPM
device.
Shock hazard —
Strictly observe the following war-
nings. Failure to do so endangers the
lives of the patient, the user and other
persons involved.
– Before use allow the OptiFlex
reach room temperature. If the
device has been transported at
temperatures below 0 °C (32 °F),
leave it to dry at room temperature
for about 2 hours, until any conden-
sation has disappeared.
– The OptiFlex
operated in dry rooms.
– When disconnecting the device from
the power line, remove the plug
®
S pro-
from the wall outlet first, before
disconnecting the cable from the
device.
– When connecting the device to
other equipment or when creating a
medical system, check that the sum
of leakage currents will not cause
any hazard. Please contact CHATTA-
NOOGA, if you have questions in
this matter.
– Do not use multiple portable socket
outlets (MPSO) to connect the device
to the power line. OptiFlex
be connected to a properly instal-
led wall outlet with a non-fused
earthed wire. Before connecting the
power cord, it must be completely
unrolled and placed such that it will
not get caught in the moving parts of
the device.
– Before cleaning and service inter-
ventions, disconnect the device
from the power line by removing
the power cord from the wall outlet.
– Liquids must not be allowed to
enter the CPM device or the
programming unit. If liquids have
entered into the devices,
OptiFlex
checked by a service technician,
before it can be reused.
13
Warning!
®
S must only be
®
S must be immediately
®
S to
®
S must