9. IEC 60601-1-2:2001
The OptiFlex
®
S device is subject to
particular precautions regarding electro-
magnetic compatibility (EMC). The
device must be installed and put into
service strictly in compliance with the
EMC directives put forth in the accom-
panying documents.
Portable and mobile RF communication
systems may affect the OptiFlex
device.
The OptiFlex
®
S device should not be
used adjacent to or stacked with other
equipment. If adjacent or stacked use is
necessary, OptiFlex
®
S should be
observed to verify normal operation in
the configuration in which it will be
used.
9.1 Electromagnetic emissions
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
OptiFlex
®
S is intended for use in the electromagnetic environment specified below. It is the
responsibility of the customer or user to ensure that the OptiFlex
environment.
Emissions test
RF emissions to CISPR 11
RF emissions to CISPR 11
Harmonic emissions to IEC
61000-3-2
Voltage fluctuations/flicker
emissions to
IEC 61000-3-3
If you detect damage or malfunctions
that may impair the safety of the patient
or of the operator, have the device
repaired before using it.
If it is necessary to replace assemblies
or cables, only the manufacturer's
original parts must be used to ensure
continued compliance with EMC
S
requirements after repair. This require-
®
ment applies to the power supply unit,
cables and cable lengths, drive unit
consisting of the motor and the control
system, the programming unit incl. the
coiled cable and the connector.
Compliance
Electromagnetic environment - guidance
Group 1
OptiFlex
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
Class B
OptiFlex
ments, including domestic and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
not applicable
not applicable
32
®
S device is used in such an
®
S uses RF energy only for its internal
®
S is suitable for use in all establish-