one of the parameter keys to select a
parameter:
• The corresponding symbol appears
on the display in a larger format.
• The set value is displayed.
• The symbol above the parameter
key appears in reverse video.
4. With the + / - keys (plus/minus) you
change the displayed value. When
you press and hold the key, the value
will change at a faster rate.
Some of the (special) functions can
only be enabled and disabled. This is
done by pressing the corresponding
parameter key or with the + / - keys.
Active parameters are identified with
a check mark in the circle next to the
symbol.
5. Having programmed all parameters,
press the STOP key to save the
values.
6. Then press the START key:
OptiFlex
®
S checks the set values,
moves to the position halfway
between the set internal and external
rotation values as well as to the set
maximum adduction value and stops.
7. Press the START key again to start
therapy.
Next the carriage will move the
maximum internal rotation value in
the synchronized mode. The car-
riage will then simultaneously move
to the maximum external rotation
position and the maximum abduction
position and subsequently to the
maximum adduction position and the
maximum internal rotation position.
After reaching this position, the cycle
restarts: maximum abduction and
maximum external rotation.
After activation of the START key in
the non-synchronized mode the
motors will perform random move-
ments, each motor reversing after
reaching the maximum values.
Note!
- Refer to sections 5.3 for a descrip-
tion of the parameters.
- To view the set parameter values,
press the corresponding parameter
key. Before, however, you have to
press the STOP key and access the
correct menu level.
- To prevent accidental changes of the
parameter settings, lock the keys by
simultaneously pressing the + (plus)
and – (minus) keys.
Press both keys again to unlock.
- Selecting the "New Patient" function
will automatically delete the data on
the patient chip card. When you have
finished programming the unit and
press the STOP key, the settings will
automatically also be saved to the
patient chip card.
- Emergency stop function:
OptiFlex
®
S will stop immediately,
when any of the keys is pressed
during therapy.
Patient treatment can be resumed
by pressing the START key. The
device will automatically change the
direction.
Patients with a programmed chip card
• First complete the mechanical adjust-
ments.
• Then insert the chip card (the patient
is not yet positioned on the CPM
device).
• Press the START key: the device will
move to the starting position as
specified by the parameters stored on
the chip card and stops.
• Position the patient on the CPM
device and press the START key to
initiate therapy.
19