Fuse replacement
Warning!
Patient hazard, equipment malfunc-
tion and damage —
The replacement of fuses must be
referred to specialists as defined in
IEC 60364 or other applicable stan-
dards (e.g. biomedical technicians,
electricians, electronics installers).
Fuses used must be T1A fuses.
• Before replacing fuses, turn off the
OptiFlex
®
S and disconnect the device
from the power line.
• Use an appropriate tool to remove
the fuse holder (8) situated between
the power switch (7) and the power
connector (6) (Fig. 1).
• Replace the fuses and reinsert the
fuse holder (Fig. 2). Check that the
fuser holder locks properly into place.
Fig. 1
Fig. 2
6.3 Transport
The following operating steps must be
completed before transporting the
OptiFlex
®
S:
1. Activate the "transport setting"
function
also 5.3) and start the OptiFlex
2. Turn off the OptiFlex
(7).
3. Remove the power cord (6) and
disconnect the motion element and
the programming unit (9).
4. Open locking screw (14) and pull out
the armrest (4) for the healthy arm.
5. Grasping the holding tube (17) of the
motion element, open locking screw
(14) and pull out the motion element.
6. Set horizontal adduction/abduction
to 0° (indexing knob 12).
7. Open the locking screw for adjust-
ment of the backrest (10), fold the
backrest to the front all the way until
flat on the seat and tighten the
screw.
8. Remove the two locking pins (30).
Pull out the legs towards the sides,
reverse and reinsert. Then put the
locking pins back in place.
28
in the menu (see
®
S.
®