Chemical
cleaning with
disinfection:
– 29 –
EQUIPMENT:
The method has been validated in Europe using:
– Aldehyde-free instrument disinfectant: Bode
®
Bomix
 plus (concentration: 2%)
Active agent basis: Quaternary ammonium
compound
– Drinking water at approx. 15 °C
– Application time: 5 minutes
PROCEDURE:
• Clean and disinfect the individual parts in a
single work step with a solution prepared ac-
cording to the manufacturer's instructions.
Info: If the recommended treatment time is
exceeded significantly, the plastic parts may
take on the smell of the disinfectant.
RINSING:
 CAUTION! Disinfectant residues can
cause allergic reactions or irritation of the mu-
cous membrane. Rinse off all parts thoroughly
in running drinking water at about 15 °C for 3
minutes.
• Dispose of the used solution (the diluted solu-
tion can be disposed of down the drain).
DRYING:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent
surface and allow them to dry completely.