Invacare® HomeFill® II
1.
Make sure the concentrator is On. Refer to the concentrator
User Manual.
The maximum oxygen concentrator flow rate for
patient use is listed in the table below. Otherwise, the
oxygen level to the compressor will be <90% and the
compressor will not fill the cylinder.
Invacare Oxygen
Concentrator
Platinum 5
Platinum 9
Perfecto
2
Patients with a higher prescribed flow should not refill a
cylinder while receiving supplemental oxygen treatment
from the oxygen concentrator.
2.
Set the concentrator flow rate to the patient to 2.5 L/min. or less
for Platinum 5 and 5 L/min. for Platinum 9 oxygen concentrator.
Refer to the concentrator User Manual.
3.
Make sure the cylinder is connected to the compressor. Refer
to 4.5.1 Connecting the Cylinder to the Compressor, page 17.
4.
Push the power switch B on the control panel A to the On
( ) position.
20
All manuals and user guides at all-guides.com
Maximum Output Flow
for Patient Use
2.5 L/min
5 L/min
3 L/min
5.
Examine the control panel. The following sequence of events
should occur:
a.
0-3 minutes - The O
2
be lit while the compressor warms up.
After three minutes - The filling (green) light D will be on
b.
while the cylinder is filling.
The full (green) light E will be on when the cylinder is
c.
finished filling. Proceed to step 6.
The O
below normal (YELLOW) light may come on
2
if the oxygen level from the concentrator has not yet
reached or drops below 90% or higher. The compressor
will begin or resume filling when the oxygen level
reaches 90%. Refer to 7.2 Cylinder Fill Times, page 27 if
light does not go Off within 10 minutes.
Refer to 7.2 Cylinder Fill Times, page 27 if the Attention
(red) light F and audible alarm are on.
6.
Remove the full cylinder. Refer to Disconnecting the Cylinder
from the Compressor, page 18.
7.
Push the power switch to the Off (O) position.
For cylinder operation and use, refer to the specific user
manual for the cylinder/regulator assembly.
below normal (YELLOW) light C will
1145804-E