Descargar Imprimir esta página

Troubleshooting; Indicator Lights - Invacare HomeFill II Manual De Instrucciones

Ocultar thumbs Ver también para HomeFill II:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 147

6 Troubleshooting

6.1 Indicator Lights

All indicators are for low priority conditions that require operator awareness only.
INDICATOR LIGHT COLOR
None
O
Below Normal -
2
YELLOW
Filling - GREEN
1145804-E
All manuals and user guides at all-guides.com
COMPRESSOR STATUS
Compressor is Off.
Compressor is On and warming up
(0-3 minutes after power is On).
O
from concentrator is < 90% (any time after
2
three minute warmup).
Compressor is On and cylinder is filling. O
from concentrator is >90% (any time after
three minute warmup).
EFFECT
O
flowing to user if concentrator is On.
2
Cylinder filling not started.
O
flowing to user if concentrator is On.
2
Cylinder filling stops, will restart when O
> 90%.
O
flowing to user if concentrator is On.
2
Cylinder is filling
2
O
flowing to user if concentrator is On.
2
Troubleshooting
2
23

Publicidad

loading

Este manual también es adecuado para:

Ioh200aw