Troubleshooting; Indicator Lights - Invacare HomeFill II IOH200AW Manual Del Usuario

Ocultar thumbs Ver también para HomeFill II IOH200AW:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 89
Invacare® HomeFill® II Oxygen System

7 Troubleshooting

7.1 Indicator Lights

Indicator Light Color
None
O
BELOW NORMAL - YELLOW
2
FILLING - GREEN
FULL - GREEN
ATTENTION - RED (Audible Alarm)
30
Compressor Status
Compressor is Off.
Compressor is On and warming up (0-3
minutes after power is On).
O
from concentrator is < 90% (any time
2
after three minute warmup).
Compressor is On and cylinder is filling.
O
from concentrator is >90% (any time
2
after three minute warmup).
Compressor is On. Cylinder is finished
filling (full) -remove cylinder (FULL light
is On in >10 mins.)
or
Cylinder is not attached properly - check
connection (FULL light is On in <10
mins.).
Compressor is On. Cylinder NOT filling.
Causes:
Full cylinder attached when unit turned
On - replace with empty cylinder
or
Effect
O
flowing to user if concentrator is on.
2
Cylinder filling not started.
O
flowing to user if concentrator is On.
2
Cylinder filling stops, will restart when
O
> 90%.
2
O
flowing to user if concentrator is On.
2
Cylinder is filling.
O
flowing to user if concentrator is On.
2
Cylinder filling stops.
Cannot fill another cylinder until
compressor is turned Off and On again.
O
flowing to user if concentrator is On.
2
Compressor does not begin to fill or
stops filling.
Unit must be turned Off and On before
filling can occur.
1195663-C