Rinse the electrode with distilled water
Press the button
Wait for the pH value on the display to stabilize; when the icon
appears, use the keys
entering the correct one (e.g. pH 6.79).
Note: Check the buffer value according to the temperature.
When the icon
appears again, press the key
the first point; the actual measured value flashes on the display and the beaker icon appears at the
bottom left with the buffer identification colour and value
value").
Rinse the electrode with distilled water
Press the button
the electrode in the next buffer (e.g. pH 4.65).
Wait for the pH value on the display to stabilize; when the icon
appears, use the keys
entering the correct one (e.g. pH 4.65).
When the icon
appears, press the key
value flashes on the display and the icons
Although the device can accept three more calibration points, stop and confirm this calibration by
pressing
.
The calibration report and the graphic representation appear on the display; press the button
or
to exit and return to measure mode. The beakers relating to the calibration are displayed at
the bottom left side, the value is preceded by the letter "U", indicating that the value has been entered
manually.
Note: If you are working with manual temperature compensation (MTC), update the value before
calibrating the instrument.

• Performing pH measurement

In measure mode, press the button
Connect the electrode to the BNC of the instrument in green colour.
Connect the temperature probe to the RCA in the green input.
If the user does not use an electrode with a built-in temperature probe or an external probe NTC 30KΩ,
it is recommended to manually update the temperature value (MTC).
Remove the electrode from its tube, rinse with distilled water and dab gently with paper towel.
Check the presence and eliminate any air bubbles in the membrane bulb by stirring vertically (as for
the clinical thermometer). If present, open the side cap.
Dip the electrode in the sample, while keeping it slightly stirred.
Consider the measurement truthful only when the stability icon
UM ReVio EN rev.0 04.06.2020
and gently dab it with paper towel.
and dip the electrode in the first buffer solution (e.g. pH 6.79).
and
to adjust the value by
and gently dab it with paper towel.
to proceed with the calibration and dip
and
to adjust the value by
U 4.65
appear at the bottom left.
and move to pH (see paragraph "Operation of the device").
21
to confirm
U 6.79
(the letter U indicates the "user
to confirm the second point; the actual measured
appears.