• Replacement of the membrane

If the membrane has ripples or irregularities, it must be replaced with a new one.
Remove the membrane cap and replace it with a new and intact one.
With the new membrane, follow the procedure of the Replacement of Electrolyte.
If, even after these procedures, the probe does not calibrate, replace the probe.

Barometric pressure

Since the measurement of the partial pressure of the dissolved oxygen is related to the barometric pressure,
this instrument is able to compensate each variation, thanks to the integrated barometric sensor.
In order to view the barometric pressure measured by the instrument, press the key
Press. The measurement is express in mbar.

Multiparametric view

Up to 6 parameters can be displayed simultaneously on the user's discretion.
Access the SETUP and and move the cursor over the icon
Press again the button
The list of all the parameters managed by the device is shown on the display. Scroll them with the
directional keys and with the button
MULTIVIEW screen. Up to 6 can be chosen.
Press
to confirm and return to measure mode. With the
button
move to the MULTIVIEW screen. For each
parameter, its relative temperature is also displayed. For graphic
reasons, there is no "smile" icon, therefore, to indicate that the
measurement has reached stability, the colour of the value on
the display changes from black to grey.

Data Logger function

This series of devices has the possibility of recording values in GLP format on the instrument's internal
memory.
The instrument can save up to 10000 data in total. Once the memory is finished, the values are NOT
overwritten. In measure mode, next to the icon
appears
It is possible to recall and consult the values on the display or download them to a PC using the
appropriate software.
If you have the possibility to work directly connected to the PC, the data are automatically saved on
the software without having memory limitations. Recommended option if you plan measurements with
durations of more than 15 hours.
Recordings can be acquired manually (MANUAL) or automatically at pre-set frequencies (SECONDS –
MINUTES-HOURS).
PC Connection : connect the USB cable inside each package to the USB port on the top panel of the
instrument and the other end to a COM port on the computer. Use the USB cable supplied with the
instrument only.
UM ReVio EN rev.0 04.06.2020
to access.
activate the flag next to those you want to display on the
, the number of data stored for that parameter
40