Capacitance
1)
RANgE
500.0 nf, 5.000 µf, 50.00 µf, 500.0 µf, 3000 µf
Resolution: 0.1 nf on 500.0 nf range
1)
Additional 50.00nf range accuracy is not specified
2)
Accuracies with film capacitor or better
3)
Specified with battery voltage above 2.8V (approximately half full battery). Accuracy
decreases gradually to 12% at low battery warning voltage of approximately 2.4V
DC Current (Clamp-on)
RANgE
0.0A to 400.0A
400A to 800A
800A to 900A
900A to 1000A
Max current: 1000A continuous
Resolution: 0.1 A on 400 A range
1)
induced error from adjacent current-carrying conductor: < 0.01A/A
2)
Zero mode must be applied to zero the unit before measurement to ensure accuracy.
ACA Current (Clamp-on)
RANgE
400.0A
400.0A
< 50A
4)
< 50A
4)
400A to 600A
400A to 600A
600A to 800A
Max current: 800A rms continuous
Resolution: 0.1 A on 400 A range
1)
induced error from adjacent current-carrying conductor: < 0.01A/A
2)
Crest factor (ACDC-100 TRMS): < 1.6 : at full scale; < 3.2 : at half scale
3)
± (4.0% + 5dgts) for ACDC- 100 TRMS
4)
Accuracy is specified at <50A in this frequency bandwidth due to limited calibrator output
capability for testing
All manuals and user guides at all-guides.com
frequency
5hz to 40hz
40hz to 200hz
200hz to 400hz
400hz to 1khz
5hz to 40hz
40hz to 100hz
5hz to 60hz
12
2) 3)
Accuracy
± (3.5% + 6dgts)
Accuracy
1) 2)
± (1.5% + 4dgts)
± (1.5% + 4dgts)
± (2.0% + 4dgts)
± (5.0% + 30dgts)
Accuracy
1) 2)
± (2.0% + 5dgts)
± (1.5% + 5dgts)
± (1.5% + 5dgts)
± (2.0% + 5dgts)
± (2.0% + 5dgts)
± (1.5% + 4dgts)
± (5.0% + 30dgts)
3)
3)