Manual de operación
00809-0809-3104
Tabla 6-2: Puntos de datos de registro horario
Punto de datos
HourlyMacro2
GUSM Rosemount modelo 3418
Contenido
bit 0
IsQFlowInvalid
bit 1
IsQBaseInvalid
bit 2
IsMassRateInvalid
bit 3
IsEnergyRateInvalid
bit 4
IsEstimatedFlowVelocityInUse
bit 5
IsTooFewOperChords
bit 6
IsFwdBaselineNotSet
bit 7
IsRevBaselineNotSet
bit 8
IsGCDataErr
bit 9
No se utiliza
bit 10
IsAcqModuleIncompatible
bit 11
IsHourlyLogFull
bit 12
IsDailyLogFull
bit 13
IsAuditLogFull
bit 14
IsAlarmLogFull
bit 15
IsSystemLogFull
bit 16
IsXdcrFiringSyncError
bit 17
IsColocMeterCommErr
bit 18
IsGCCommErr
bit 19
IsGCAlarmPresent
bit 20
IsElecVoltOutOfRange
bit 21
IsElecTempOutOfRange
bit 22
DidCnfgChksumChg
bit 23
DidPowerFail
bit 24
IsAcqModuleError
bit 25
IsAcqMode
bit 26
DidColdStart
bit 27
IsCorePresent
bit 28
WatchDogReset
bit 29
DI1
bit 30
IsWarmStartReq
bit 31
IsClkInvalid
(continuación)
Directorio
Abril 2022
Acción de re-
gistro
MACRO
95