Turbidity Measurement Procedure; Measurement Techniques - HACH LANGE 2100N Manual Básico Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75

Turbidity measurement procedure

1. Rinse a clean,
2. Clean the sample
empty sample cell two
cells with a soft, lint-free
times with the solution
cloth to remove water
to be measured and
spots and fingerprints.
drain to waste. Fill to
the line (about 30 mL)
with sample and
immediately put the cap
on the sample cell.
7. Read and record the
value when stable.
Note: To send (via
RS232) a measurement
record, push PRINT.

Measurement techniques

Measurements may be made with different operation mode settings and
optional accessories.
3. Apply a small bead
4. Use the oiling cloth
of silicone oil from the
provided to apply the oil
top to the bottom of the
equally to the surface of
sample cells.
the sample cells.
Remove the excess oil.
Make sure that the
sample cells are almost
dry.
Calibrate the instrument whenever the sample cell pathlength is
changed.
5. Gently and slowly
6. Put the sample cell
invert the sample cell to
in the sample cell
fully mix the sample. Be
holder with the triangle
careful not to add air
on the sample cell
bubbles.
aligned with the
reference mark on the
sample cell holder.
Close the cover.
English 13

Publicidad

Tabla de contenido
loading

Tabla de contenido