Specifications
Specifications are subject to change without notice.
Specification
Details
Measurement method
Nephelometric
Regulatory
Meets EPA Method 180.1
ASTM D7315 - Standard Test Method for
Determination of Turbidity Above 1 Turbidity Unit (TU)
in Static Mode
ASTM D6655 - Standard Test Method for
Determination of Turbidity Below 5 NTU in Static
Mode
Light source
Tungsten filament lamp
Measurement modes
NTU, NEPH (Nephelo) and EBC
Standard operation
on page 9
Maintenance
on page 14
Troubleshooting
on page 17
Specification
Details
Range
NTU (Ratio on, manual range): 0-0.999, 0-9.99,
0-99.9, 0-4000
NTU (Ratio on, auto range): 0-4000 auto decimal
Nephelo (Ratio on, manual range): 0-9.99, 0-99.9,
0-26,800
Nephelo (Ratio on, auto range): 0-26,800 auto
decimal
Nephelo (Ratio off): 0-268
0-99.9, 0-980
EBC (Ratio on, auto range): 0-980 auto decimal
1, 2
Accuracy
Ratio on: ±2% of reading plus 0.01 NTU from
0-1000 NTU, ±5% of reading from 1000-4000 NTU
based on formazin primary standard
Ratio off: ±2% of reading plus 0.01 NTU from
0-40 NTU (under reference conditions
Resolution
Turbidity: 0.001 NTU/EBC, Nephelo (on lowest range)
Repeatability
±1% of reading or 0.01 NTU, whichever is greater
(under reference conditions)
Response time
Signal averaging off: 6.8 seconds
Signal averaging on: 14 seconds (when
10 measurements are used to calculate the average)
Stabilization time
Ratio on: 30 minutes after start-up
Ratio off: 60 minutes after start-up
Reading modes
Manual or auto range, signal averaging on or off,
Ratio on or off
Power requirement
115-230 VAC, 50/60 Hz (automatic power selection)
28 W maximum