2.1 Explanation of Symbols MIGHTY + MEDIC
Please follow these instructions carefully!
(according to IEC 60601)
ON/OFF Switch
Symbol for serial number - followed by the serial number of the
medical product
Symbol for the manufacturer - the manufacturer's name and
address are next to the symbol
The medical product was introduced after 13 August 2005. The
product may not be disposed of in normal household waste. The
X on the trash can indicates the necessity of disposing of this
product separately. The Vaporizer contains a built-in rechargeab-
le lithium-ion battery that must be discharged before disposal.
The Vaporizer is protected by an angle of 15° against dropping
IP 22
water (IP 22). The Vaporizer should nevertheless be protected
against moisture and rain. Do not use in the bathroom or above
water.
Safety Category II device
EU conformity symbol: A 4-digit number after the CE marking
0123
indicates that a notified body is involved in the conformity
assessment procedure.
Symbol for order number - followed by the order number of the
corresponding medical product (or part)
Safety tested and production monitored by TUEV SUED accor-
ding to IEC 60601-1, CAN/CSA C22.2 No. 60601-1 and ANSI/
AAMIES 60601-1.
Caution! Hot surfaces!
2 F O R Y O U R S A F E T Y
46