EN
SAFIRA Driver Indicator Lights:
8.
On the top edge of the SAFIRA Driver (refer to Diagram #21) colour indicator lights
are displayed represen ng various func ons in ac on or func ons that require
addi onal ac on. The indicators are as follows:
LED Colour
Red
Red
Green
Yellow
No Ac ve
Light
Green =
Infusion
Diagram #21: Top edge of driver, illustrated with all indicators.
MP-0007-3001 v4.0
State
Flashing
(Infusion has
stopped)
Steady
(End of Driver Life)
Steady
(Infusion Func on
Ac ve)
Steady
(Aspira on Func on
Ac ve)
N/A
Yellow =
Aspira on
Meaning
Warning:
Immediate
response by
operator is
required
2. Reset SAFIRA (see steps #10 and
Warning:
Immediate
response by
operator is
required.
Normal: System
opera ng as
designed indica ng
infusion ac vity
Cau on: Aspira on
occurring - system
opera ng as
intended
Normal: System is
ready for infusion
or aspira on
Flashing Red =
Infusion Stopped
Operator Response
Operator: Infusion has stopped;
1. Verify needle patency and/or
reposi on needle to allow low
pressure infusion
#11 for details)
Operator: Immediately disconnect
the syringe and either replace
Driver or con nue the procedure
manually.
Operator: Con nue infusion
ac vity un l medical result is
achieved.
Operator: Follow Standard
Precau ons during use of
aspira on func on
Operator: N/A
Steady Red =
End of Driver Life
Page 12 of 64