Electromagne c Immunity (con nued):
a
Field strengths from fixed transmi ers, such as base sta ons for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theore cally with accuracy. To assess the electromagne c
environment due to fixed RF transmi ers, an electromagne c site survey should be
considered. If the measured field strength in the loca on in which SAFIRA is used exceeds
the applicable RF compliance level above, SAFIRA should be observed to verify normal oper-
a on. If abnormal performance is observed, addi onal measures may be necessary, such as
re-orienta ng or reloca ng SAFIRA.
b
Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guide to SAFIRA References:
900016
900058
900029
900044
900059
MP-0007-3001 v4.0
End of Sec on
———————————————————
(Component Level—stands for the SAFIRA Luer Syringe component)
(Component Level—stands for the SAFIRA NRFit Syringe component)
(Component Level—stands for the SAFIRA Driver component)
(Component Level—stands for the SAFIRA Foot Pedal Operator
Component)
(Component Level—stands for the SAFIRA Palm Operator
Component)
End of Sec on
———————————————————
Page 18 of 64