Electromagne c Emission:
Guidance and Manufacture's Declara on: Electromagne c Emission
SAFIRA is intended for use in the electromagne c environment specified below.
The customer and/or user of SAFIRA should assure that it is used in such an environment.
Emission Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Electromagne c Immunity
Guidance and Manufacture's Declara on: Electromagne c Immunity
SAFIRA is intended for use in the electromagne c environment specified below.
The customer and/or user of SAFIRA should assure that it is used in such an environment.
Immunity Test
Electrosta c
Discharge (ESD)
IEC 61000-4-2
Power Frequency
(50Hz/60Hz)
Magne c Field
IEC 61000-4-8
Note: U
MP-0007-3001 v4.0
Compliance
Group 1
SAFIRA uses RF energy only for its internal func on.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
Group A
SAFIRA is suitable for use in all establishments, other
IEC 60601 Test
Compliance
Level
±8 kV contact
±8 kV contact
±2, 4, 8, 15 kV
±2, 4, 8, 15 kV
air
30 A/m
is the a.c. mains voltage prior to applica on of the test level.
T
Electromagne c Environment: Guidance
equipment.
than domes c and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domes c purposes.
Electromagne c Environment:
Level
Floors should be wood, concrete or
ceramic le. If floors are covered with
air
synthe c material, the rela ve
humidity should be at least 30%.
30 A/m
Power frequency magne c fields
should be at levels characteris c of a
typical loca on in a typical
commercial or hospital environment.
Guidance