General Specifications; Electrical Specifications - Klein Tools CL220 Manual De Instrucciones

Ocultar thumbs Ver también para CL220:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ENGLISH

GENERAL SPECIFICATIONS

Klein Tools CL220 is an automatically ranging, true root-mean-
square (TRMS) digital clamp-meter that measures AC current via the
clamp, AC/DC voltage, resistance and continuity via test-leads, and
temperature via a thermocouple probe.
• Operating Altitude: 6562 ft. (2000m)
• Relative Humidity: <95% non-condensing
• Operating Temp: 32° to 122°F (0° to 50°C)
• Storage Temp: 14° to 122°F (-10° to 50°C)
• Accuracy: Values stated at 65° to 83°F (18° to 28°C)
• Temp Coefficient: 0.1 x (Quoted Accuracy) per °C above
28°C or below 18°C, corrections are required when ambient
working temp is outside of Accuracy Temp range
• Dimensions: 8.46" x 3.54" x 1.50" (215 x 90 x 38 mm)
• Weight: 11.04 oz. (313 g) including batteries
• Calibration: Accurate for one year
• Auto Power-Off (APO): After approx. 10 minutes of inactivity
• Standards: IEC EN 61010-1, 61010-2-032, 61010-2-033.
IEC EN 61326-1, 61326-2-2.
Conforms to UL STD.61010-1,
61010-2-032,61010-2-033;
Certified to CSA STD.C22.2 NO. 61010-1,
61010-2-032,61010-2-033.
• Pollution degree: 2
• Accuracy: ± (% of reading + # of least significant digits)
• Drop Protection: 6.6 ft. (2m)
• Safety Rating: CATIII 600V, Class 2, Double insulation
• Electromagnetic Environment: IEC EN 61326-1. This
equipment meets requirements for use in basic and controlled
electromagnetic environments like residential properties,
business premises, and light-industrial locations.
Specifications subject to change.
2

ELECTRICAL SPECIFICATIONS

Function
Range
200.0mV
2.000V
AC Voltage
20.00V
(V AC)
200.0V
600V
200.0mV
2.000V
DC Voltage
20.00V
(V DC)
200.0V
600V
Input Impedance: 10MΩ
Frequency Range: 45 to 400Hz
Maximum Input: 600V AC RMS or 600V DC
2.000A
20.00A
AC Current
(A AC)
200.0A
400A
Frequency Range: 50 to 60Hz
200.0Ω
2.000KΩ
20.00kΩ
Resistance
200.0kΩ
2.000MΩ
20.00MΩ
Maximum Input: 600V AC RMS or 600V DC
-40° to 10°F
Temperature
°F
11° to 1832°F
-40° to -12°C
Temperature
°C
-11° to -1000°C
OTHER MEASUREMENT APPLICATIONS
Maximum Input: 600V AC RMS or 600V DC
• Continuity Check: Audible signal <10Ω, max current 1.5mA
• Sampling Frequency: 3 samples per second
• Overload: "OL" indicated on display
• Polarity: "-" on display indicates negative polarity
• Display: 3 ½ digit, 2000 Count LCD
Resolution
Accuracy
0.1mV
±(2.5% + 10 digits)
1mV
10mV
±(2.0% + 5 digits)
100mV
1V
0.1mV
±(1.0% + 8 digits)
1mV
10mV
±(1.0% + 3 digits)
100mV
1V
1mA
±(2.5% + 30 digits)
10mA
100mA
±(2.0% + 10 digits)
1A
0.1Ω
±(1.2% + 5 digits)
10Ω
±(1.2% + 3 digits)
100Ω
1kΩ
10kΩ
±(2.0% + 5 digits)
±(1.2% + 7 digits)
1°F
±(1.2% + 6 digits)
±(1.2% + 4 digits)
1°C
±(1.2% + 3 digits)
3

Publicidad

Tabla de contenido
loading

Tabla de contenido