Section 5- Measurement Validation
The Pulse oximeter accuracy has been validated in human studies against arterial
blood sample reference measured with a CO-Oximeter. In a controlled desaturation
study, healthy adult volunteers with saturation levels between 70% and 100% SaO2
were studied.
Subject Demographics
The population characteristics for those studies as follow table 5
Table 5—PULSE OXIMETER Clinical study Subject Demographics Record.
Subject
Gender Age
#
1#
M
2#
M
3#
F
4#
M
5#
F
6#
F
7#
F
8#
M
9#
F
10#
M
11#
F
12#
M
ARMS Results
The final analysis was performed on 241 data points collected across 12 subjects.
The SpO2 accuracy performance of each pulse oximeter and sensor combination is
identified below
Where:
ARMS is the accuracy root mean square.
SpO2 is the test pulse oximeter readings during sample i.
Sao2 is the Average Reference CO-Oximeter functional oxygen saturation reading
during sample i.
n is the number of points.
Height
Weight
(cm)
(kg)
31
160
70
24
165
55
22
160
45
29
175
60
22
160
49
19
160
45
21
162
54
34
192
102
27
178
58
23
178
78
24
174
80
26
169
65
English 21
Skin Tone
Light
Light
Light
Medium Dark
Light
Light
Light(White)
Light(White)
Light(White)
Dark dark
Dark dark
Dark dark
Remark
Asian (Chinese)
Asian (Chinese)
Asian (Chinese)
Asian (Chinese)
Asian (Chinese)
Asian (Chinese)
Caucasian
Caucasian
Caucasian
African
African
African