13 Appendix
Command
r/w
Authorization Reply
bsz x
r
r-all
Settings
lang
r/w
r-all
w-Setter
EWR 2 Net
copid x
r/w
r-all
w-Setter
copbd x
r/w
r-all
w-Setter
ip w x y z
r/w
r-all
w-Setter
nmask w x y z
r/w
r-all
w-Setter
Gateway
r/w
r-all
w x y z
w-Setter
dhcpIP x
r/w
r-all
w-Setter
ga x
r/w
r-all
w-Setter
gf x y
r/w
r-all
w-Setter
gn x y
r/w
r-all
w-Setter
Tab. 35
Calibration/parameters
EN - 22
Description
bsz x yyyy
Operating hours counter (all times in sec.)
x = Counter type:
0: Number of power cycles (on/off switch cycles)
1: Time in undervoltage (< 24 V − 10%)
2: Time in normal voltage
3: Time in overvoltage (> 24 V + 10%)
4: Time with flow regulation active
5: Time with pressure regulation active
6: Gas consumption amount in L
10..19: Error memory 0..9
7.3.2 on page EN-12
20..29: Time stamp for error number 0..9
lang x
Setting for display menu language
0 English
1 German
2 Spanish
copid x
Can Open Node ID
x = Can Open Node ID
copbd x
Can Open baud rate (speed)
x = 0, 125 kBit/s
x = 1, 250 kBit/s
x = 2, 500 kBit/s
x = 3, 1000 kBit/s
ip w x y z
IP address for TCP server
A restart is required to adopt the IP
nmask w x y z Netmask for TCP server
A restart is required to adopt the IP
Gateway
Gateway for TCP server
w x y z
A restart is required to adopt the IP
dhcpIP x
Change-over switch DHCP/static IP
x = 1 DHCP
x = 0 static IP
ga x
Selection of gas type to be used
x = Index of current gas
0...9 set gas
gf x y
Configuring gas factors for the gas type (ga)
x = gas index
y = factor in 1/10% (1000 = 1.0)
gn x y
Configuring the gas name
x = gas index
y = gas name as string (max. 10 characters)
EWR 2 / EWR 2 Net
Value range
Long 4Byte
0...0xFFFFFFFF
Seconds
resolution up to
136.2 years
x = 0 to 2
x = 0 to 255
x = 0 to 3
w = 0 to 255
x = 0 to 255
y = 0 to 255
z = 0 to 255
w = 0 to 255
x = 0 to 255
y = 0 to 255
z = 0 to 255
w = 0 to 255
x = 0 to 255
y = 0 to 255
z = 0 to 255
x = 0,1
x = 0 to 9
x = 0 to 9
y = 500 to 1500
x = 0 to 9
y = string
BAL.0598 • 2018-04-09