Descargar Imprimir esta página

Amprobe SOLAR-600 Manual De Uso página 25

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

SPECIFICATION
Electrical Specifications
DC Voltage Measurement
Range
0 ~ 10 V
10 ~ 60 V
Vopen: open circuit voltage of solar cell or module.
If users use alligator clips to measure voltage only (I+ clip is not connected),
clips (V- and I-) must be shorted together. Thus, 4-wire measurement is
converted to 2-wire measurement.
DC Current Measurement
Range
0.01 ~ 10 A
10 ~ 12 A
Ishort: short circuit current of solar cell or module.
Circuit resistance is compensated in the AUTO SCAN. Ishort is measured at
zero circuit resistance. Circuit resistance is not compensated in the Manual
Scan or Single Point Test.
DC Current Simulation
Range
0.01 ~ 10 A
10 ~ 12 A
If current is greater than 12A, test (Auto-Scan, Scan, or Test) can not be
performed.
(23°C ±5°C, Four-wire Measurement)
Resolution
0.001 V
0.01 V
Resolution
1 mA
10 mA
Resolution
1 mA
10 mA
18
Accuracy
± 1 % ± (1 % of Vopen ± 0.1 V)
± 1 % ± (1 % of Vopen ± 0.1 V)
Accuracy
± 1 % ± (1 % of Ishort ± 9 mA)
± 1 % ± (1 % of Ishort ± 0.09A)
Accuracy
± 1 % ± 9 mA
± 1 % ± 0.09A

Publicidad

loading