Table 4. Problems, possible causes, and possible solutions (continued)
Problem: Telemetry Status light flashes, but no beep sounds.
Possible cause: The speaker has failed or the area is too noisy to hear the beep.
Possible solution: Try using the Patient Assistant in a quieter area; if you still
do not hear a beep, contact your doctor or clinic to replace the Patient
Assistant.
Problem: Telemetry Status light does not flash and no beep sounds.
Possible cause: Batteries are inserted incorrectly. Possible solution: Reinstall
the batteries (see page 30).
Possible cause: Batteries are depleted. Possible solution: Replace the batteries.
Possible cause: Wrong type of batteries are inserted. Possible solution: Verify
that batteries are N-size, 1.5 V.
Possible cause: Component failure. Possible solution: Contact your doctor or
clinic to replace the Patient Assistant.