.- May it be necessary, change the membrane and the Chlorine electrode cell every six months of functioning. To do this, disconnect
the Chlorine cell (15) from the PW Basic and take it out from the analysis chamber. Afterwards, unscrew the protector, leaving it
outside the electrode. Take the electrolyte from inside the protector and then unscrew the protector closure, leaving the membrane
free. Substitute the membrane for a new one, wind the closure to the protector and fill it with the electrolyte, trying not to leave any
air inside. Finally, clean the electrode with the electrolyte and wind the electrode to the protector.
.- Put the Chlorine cell incide the analysis chamber and connect the cell to the PW Top. Whenever you do this, you will have to adjust
the Chlorine cell afterwards.
.- In case the swimming pool is not working (end of season, repairs, etc.) the PW Top Inorganic Chlorine will be switched
off and the pH probe and Chlorine cell will be taken out from the chlorine chamber. The pH electrode and the chlorine cell must be
cleaned with water, putting in the extremes their plastic protectors and keeping them in their boxes. The holes where the pH
electrode and the Chlorine cell were placed, will be closed with two plastic lids to avoid getting dirty. Protect the PW Basic
equipment from the dirt and the humidity.
It is advisable to revise PoolWatch Top Inorganic Chlorine once a year by a specialist technician.
14. INORGANIC CHLORINE CELL WORK INSTRUCTIONS