Upper and lower illumination. LED illumination with adjustable brightness. The illumination system
can be powered by AA batteries or by A/C. The upper illumination is used to observe opaque objects
whereas the lower one allows observing transparent objects. Use the both illuminations to study
semi-transparent objects.
A/C adapter socket. Is used to connect the microscope to A/C power.
Using the microscope
Getting started
• Unpack the microscope and make sure all parts are available.
• Move the stage to the lowermost position using the focusing knob.
• Make sure the batteries are correctly installed in the battery compartment; insert new batteries
if required. Or, you can connect the microscope to A/C power using the A/C adapter.
• Insert the eyepiece into the eyepiece tube.
• Slowly adjust the illumination brightness, from dark to light.
Focusing
• Place a specimen on the stage and fix it with the holders.
• Select the 4x objective rotating the revolving nosepiece.
• Move the specimen to place its thickest part exactly under the objective.
• Rotate the focusing knob to slowly raise the stage until the objective is close to the specimen; keep
checking the distance between the objective and the object to avoid their contact.
CAUTION: The objective should not touch the specimen, otherwise the objective or/and the
specimen might be damaged.
• Look through the installed eyepiece and lower the stage slowly rotating the focusing knob until
you see the specimen image.
• Such adjustment protects the frontal lens from contacting the object when you use objectives of
other magnifications; though, slight refocusing might be required.
• If the image is too bright, rotate the diaphragm disk until the passing light ray is reduced to a
comfortable brightness level. If the image is too dark, select a larger aperture to increase the light ray.
Selecting the objective
Start your observations with the lowest magnification objective and select a specimen segment
for detailed research. Then move the specimen to center the selected segment in the field of view,
to make sure it keeps centered when the objective is changed to a more powerful one. Once the
segment is selected, you should center its image in the microscope's field of view as precisely as
possible. Otherwise, the desired segment might fail to center in the field of view of the higher power
objective. Now you can switch to a more powerful objective by rotating the revolving nosepiece. Adjust
the image focus if required.
Digital camera
The digital camera is installed in the eyepiece tube in place of the eyepiece. It allows to observe
specimens in fine detail and true colors on your PC monitor and save images on the hard drive.
Levenhuk sofmware allows viewing and editing the object images.
Megapixels
Video recording option
Mounting location
Image format
Exposure
White balance
Exposure control
Sofmware
Sofmware image control
Port
eyepiece tube (replaces the eyepiece)
*.bmp, *.jpg, *.jpeg, *.png, *.tif, *.tiff, *.gif, *.psd, *.ico, *.emf, etc.
image size, brightness, exposure control
USB 2.0, 480Mb/s
2.0
+
ERS
auto/manual
auto/manual
Levenhuk
7