Typical Measurements and Accuracy (cont'd)
Current Measurement
(cont'd)
AC and DC Clamp
AC Current (CM-960 and CMI-600 only)
Range
Accuracy Range
0.00 A to 8.99 A
60.00 A
9.00 A to 60.00 A
600.0 A
60.0 A to 600.0 A
* Maximum crest factor < 2:1 at full scale and < 4:1 at half scale.
AC Current (CM-660, CM-860, and CMH-600 only)
Range
Accuracy Range
0.00 A to 0.19 A
60.00 A
0.20 A to 5.99 A
6.00 A to 60.00 A
600.0 A
60.0 A to 600.0 A
* For any measurement, when the conductor is not positioned on the center of the
jaws, add 2.0% to the accuracy for position error.
DC Current (CM-960 and CMI-600 only)
Range
60.00 A
600.0 A
AC + DC Current (CM-960 and CMI-600 only)
Range
Accuracy Range
0.00 A to 8.99 A
60.00 A
9.00 A to 60.00 A
600.0 A
60.0 A to 600.0 A
† Induced error for adjacent current carrying conductors is < 0.01 A/A.
Note: Use Δ if there is an offset before the measurement is taken.
CM-660 • CM-860 • CM-960 • CMH-600 • CMI-600
Hz
A
Accuracy at
50 Hz to 100 Hz*†
± (1.8% + 0.15 A)
± (1.8% + 0.05 A)
± (1.8% + 0.5 A)
Accuracy at
50 Hz to 100 Hz*†
unspecified
± (1.8% + 0.15 A)
± (1.8% + 0.05 A)
± (1.8% + 0.5 A)
Accuracy Range
0.00 A to 8.99 A
9.00 A to 60.00 A
60.0 A to 600.0 A
Accuracy at
50 Hz to 100 Hz†
± (2.2% + 0.15 A)
± (2.2% + 0.05 A)
± (2.2% + 0.5 A)
Accuracy at
100 Hz to 400 Hz*†
± (2.0% + 0.15 A)
± (2.0% + 0.05 A)
± (2.0% + 0.5 A)
Accuracy at
100 Hz to 400 Hz*†
unspecified
± (2.0% + 0.15 A)
± (2.0% + 0.05 A)
± (2.0% + 0.5 A)
Accuracy†
± (2.0% + 0.15 A)
± (2.0% + 0.05 A)
± (2.0% + 0.5 A)
Accuracy at
100 Hz to 400 Hz†
± (2.7% + 0.17 A)
± (2.7% + 0.07 A)
± (2.7% + 0.7 A)
15