Cleaning Recommendation For The Non-Sterilisable M.blue Plus Instruments; Cleaning And Disinfection Of The M.blue Check-Mate - MIETHKE M.blue plus Instruments Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23
M.blue plus Instruments
This is a non-binding recommendation.
The physician must decide in each indi-
vidual case.
CLEANING RECOMMENDATION
FOR THE NON-STERILISABLE
M.blue plus Instruments
NOTE
M.blue plus Instruments are made
from thermolabile components
that are sensitive to heat and
moisture as well as chemically
reactive. Do not immerse M.blue
plus Instruments in cleaning solu-
tions or allow liquids to enter the
housing, as this could impair the
functioning as a result of mois-
ture, corrosion and impurities.
Surface impurities on the M.blue plus
Instruments should always be removed
immediately after use by wiping with
alcohol-based detergents (minimum 75
% alcohol).
The exposure time should be at least
60 seconds and should be adapted to
the degree of contamination. The instru-
ments should then be wiped with a dry
cloth.
The following cleaning methods are
not suitable for preparation of the
M.blue plus Instruments (apart from
M.blue Check-mate): Irradiation, ultra-
sound, sterilisation, mechanical prepa-
ration, immersion in cleaning fluids.
CLEANING AND DISINFECTION
OF THE M.blue Check-mate
Avoid damaging the product due to
unsuitable detergents / disinfectants
and/or too high temperatures!
Use detergents / disinfectants that
are approved for surgical steels,
according to the manufacturer's
instructions.
Observe specifications regarding
concentration, temperature and
contact time.
INSTRUCTIONS FOR USE | GB
Do not exceed the maximum per-
missible cleaning temperature of
55°C.
Perform ultrasonic cleaning:
as effective mechanical sup-
port for manual cleaning/disin-
fection
for precleaning products with
dried residues before the
mechanical cleaning / disin-
fection
as integrated mechanical sup-
port during mechanical clean-
ing / disinfection
for subsequently cleaning
products with residues that
have not been removed after
mechanical cleaning / disin-
fection
If the instruments cannot be fas-
tened securely or safely for clean-
ing in machines or on storage aids,
clean and disinfect the instruments
mechanically.
Manual cleaning / disinfection
Check visible surfaces for residues
after manual cleaning / disinfec-
tion.
If necessary, repeat the cleaning
process.
Mechanical cleaning / disinfection
Place product on cleaning-compatible
screen basket.
17

Publicidad

Tabla de contenido
loading

Tabla de contenido