Descargar Imprimir esta página

Barloworld Scientific MCS-102L Manual Del Usuario página 21

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

M i c r o C a r r i e r S t i r r e r
MCS-101L
Remove the two screws which holds the location spacer and
magnet carrier to the pulley.
Fit the appropriate magnet spacer to move the magnet position,
near the spindle for 3 litre vessels, away from the spindle for 5 litre
vessels. (The spacer and location boss for 3 litre vessels are
screwed inside the base of the platform).
Use the two screws to replace the spacer and carrier on the pulley.
Place the magnets so that they touch the spacer.
Note: Magnets must be mounted with opposite poles upwards
(opposite poles attract).
Spacer
Magnets
Other MCS Units
Remove the screw holding the
magnet to the pulley.
Position the magnet on the appropriate
peg as shown in the table.
Position
Vessel Size
Peg A
500 ml
Peg B
1 litre
Peg C
125/250 ml
Magnets must be mounted with their
blue spots (south poles) leading.
Replace the screw to hold the
magnet to the pulley.
O P E R A T O R ' S M A N U A L
Screws
Carrier
P A G E 19