STARTER 300C Portable Conductivity Meter
3.3
Set calibration standard
After the MTC temperature value setting, the meter goes to setup the calibration standard.
Use
or
buttons to select the calibration standard.
Press button-
Continue with calibration standard setting by following instructions or press button-
back to the measurement screen.
3.4
Set temperature correction coefficient
After the calibration standard setup, the meter enters the temperature correction coefficient
setting. Use
or
to 10.00 %/°C). Press button-
setting or press button-
Note:
If you set the temperature correction coefficient value as 0 %/°C, it means there will be NO
Temperature Compensation for the conductivity measurement. The meter displays the real
conductivity value at current temperature.
3.5
Set TDS factor
After selecting the reference temperature, the current TDS factor blinks. Use
buttons to increase or decrease the value. Press
automatically to the measurement screen.
to confirm your setting.
buttons to increase or decrease the temperature correction value (0.00
to confirm your setting. Continue with reference temperature
to go back to the measurement screen.
to confirm your setting. The meter will exit
EN
7
-
to go
or