a) Remove the probe from the probe holder, without disconnecting it.
b) Remove the probe holder nut and replace it with the stopper supplied.
If the probe is not already installed :
Connect the probe to the electronics unit.
4) Turn on the electronics unit.
5) Go to the «
pH Regulation - Calibration
6) Navigate through the menus following the instructions below :
pH Regulation
Calibration
OK
pH Calibration
Solution
OK
pH Calibration
In Progress
(Wait a few seconds)
pH Calibration
Solution
OK
pH Calibration
In Progress
(Wait a few seconds)
pH Calibration
Success
or
pH Calibration
Failed
3.6.15. Bluetooth communication
Menu
Setting
Mode
Communication
Pairing
Bluetooth
Reset
→ During an update of the software of the electronics unit carried out using Bluetooth, the 2 LEDs (red and green)
flash alternately.
» menu.
→ Insert the probe into the pH 7 calibration solution, then wait a few minutes.
7.0
→ Do not touch the probe.
→ a) Rinse the probe under running water, then leave to drip-dry it without wiping it.
10.0
b) Insert the probe into the pH 10 solution, then wait a few minutes.
→ Do not touch the probe.
→ a) Rinse the probe under running water, then leave to drip-dry it without wiping it.
b) Install the probe into the probe holder.
→ Carry out the navigation again with the above instructions, several times if necessary.
If calibration still fails, replace the probe and carry out another calibration.
Activation/deactivation
communication.
• Detection of connectible devices near
the electronics unit (within 60 seconds).
• Networking of the electronics unit and
connected devices.
Removal of the network connecting the
electronics unit to the connected devices.
Function
of
Bluetooth
Possible settings
Default setting
• ON
• OFF
-
ON