For transmitters rated at a maximum output power not listed above, the recommended separation dis-
tance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
• Guidance and manufacturer's declaration – electromagnetic
immunity
The device is intended for use in the electromagnetic environments
listed below, and should only be used in such environments:
Immunity
IEC 60601
Compliance
test
test level
level
Conducted
3 Vrms
Not Ap-
RF
150 kHz to
plicable
IEC 61000-
80 MHz
4-6
Radiated
3 V/m
10 V/m
RF
80 MHz to
IEC 61000-
2.5 GHz
4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
device is used exceeds the applicable RF compliance level above, the device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
3.8
7.3
12
23
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
80 MHz to 800 MHz
d = 2 .3
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compli-
ance level in each frequency range.b
Interference may occur in the vicinity of equipment marked
with the following symbol:
27
TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
POSSIBLE
SYMPTOMS
CAUSES
Unit does not
Batteries have
turn on when the
run down.
START/STOP
Battery polarities
button is pushed.
have been positioned
incorrectly.
EE measurement
Cuff has been placed
error symbol shown
incorrectly.
on display or the
Did you talk or
blood pressure
move during
value is displayed
excessively low
measurement?
Shaking of the wrist
(or high).
with the cuff on.
E1 error symbol
Air circuit
shown on display
abnormality.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Error determining
shown on display
measurement data.
EP error symbol
System error.
shown on display
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
CORRECTION
Replace them
with two new "AAA"
alkaline batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff properly
so that it is positioned
correctly. Measure again.
Keep wrist steady during
measurement. Measure
again.
Measure again.
Switch the unit off, then
measure again.
Measure again.
Measure again.
28