Maintenance Of The Sensor - ProMinent DULCOTEST PER Serie Instrucciones De Servicio

Sensor de peróxido de hidrógeno
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Maintenance of the sensor

Maintenance frequency
Maintenance operation
Cleaning the diaphragm Do not unscrew the diaphragm cap!
38
9
Maintenance of the sensor
IMPORTANT
• Maintain the sensor regularly in order to avoid over metering
in a control system resulting from an incorrect measured
value!
• Observe the relevant national regulations in force for fre-
quency of maintenance!
• Do not touch the electrodes or bring them into contact
with greasy substances!
• Do not unscrew the diaphragm cap when cleaning the dia-
phragm!
Figures based on
experience for:
Other applications: depending on operating conditions
Check the sensor regularly for dirt, deposits and air bubbles!
Avoid, as far as is possible, contamination of the diaphragm with
particles, deposits/sediments, etc. Eliminate air bubbles by in-
creasing the flow.
Check the sensor display value at the control equipment
regularly using suitable reference methods (Photometry,
Chap. 8.2).
If necessary, re-calibrate the sensor (see section 8.2 Calibration).
If calibration is no longer possible, clean or replace the dia-
phragm cap and then repeat calibration (see sections 6,
Assembly, 8.1 Running-in time and 8.2 Calibration).
Wipe the diaphragm with a damp cloth.
media with minor dirt contamination: 1 month
ProMinent
®

Publicidad

Tabla de contenido
loading

Tabla de contenido