ProMinent CTE 1-mA Instrucciones De Servicio

ProMinent CTE 1-mA Instrucciones De Servicio

Sensor de cloro total
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Enlaces rápidos

Betriebsanleitung / Operating Instructions
Mode d'emploi / Instrucciones de servicio
®
DULCOTEST
CTE
Sensor für Gesamtchlor
Sensor for total chlorine
Sonde de chlore total
Sensor de cloro total
Typ / Type / Type / Tipo CTE 1-mA
CTE 1-DMT
DE/EN/FR/ES
Teile Nr./Part No.: 987371 ProMinent Dosiertechnik GmbH · 69123 Heidelberg · Germany BA DT 106 12/09 DE/EN/FR/ES

Publicidad

Tabla de contenido
loading

Resumen de contenidos para ProMinent CTE 1-mA

  • Página 1 Sensor for total chlorine Sonde de chlore total Sensor de cloro total Typ / Type / Type / Tipo CTE 1-mA CTE 1-DMT DE/EN/FR/ES Teile Nr./Part No.: 987371 ProMinent Dosiertechnik GmbH · 69123 Heidelberg · Germany BA DT 106 12/09 DE/EN/FR/ES...
  • Página 2 Mode d’emploi en français de la page 47 à la page 68 Instrucciones de servicio en español de página 69 hasta página 91 Technische Änderungen vorbehalten. Subject to technical changes. Sous réserve de modifications techniques. Reservadas modificaciones técnicas. ® ProMinent...
  • Página 3: Tabla De Contenido

    In Betrieb nehmen ............8.1 Einlaufzeit .............. 8.2 Kalibrieren ............. Sensor warten ............... Fehler beheben ............. Außer Betrieb nehmen ..........Reparieren ..............Entsorgen ..............Technische Daten ............Bestellhinweise ............. Eingehaltene Richtlinien und Normen ......Klemmenanschlussplan DMT Chlor ......® ProMinent...
  • Página 4: Hinweise Zum Lesen Der Betriebsanleitung

    Diese Betriebsanleitung enthält die Produktbeschreibung in Fließtext sowie • Aufzählungen, Handlungsanweisungen und Sicherheitshinweise mit Symbolen gekennzeichnet: VORSICHT Bei Nichteinhalten der Sicherheitshinweise besteht die Gefahr leichter Körperverletzung und Sachbeschädigung. ACHTUNG Bei Nichteinhalten der Sicherheitshinweise besteht die Gefahr einer Sachbeschädigung. HINWEISE Arbeitshinweise ® ProMinent...
  • Página 5: Sicherheit

    • Überprüfen Sie die Lieferung auf Vollständigkeit anhand Ihrer Bestellung und der Lieferpapiere. Lieferumfang • 1 Sensor CTE 1 • 1 Flasche mit Elektrolyt (50 ml) und Tülle • 1 Ersatzmembrankappe • 1 Betriebsanleitung • 1 kleiner Schraubendreher ® ProMinent...
  • Página 6: Lagern Und Transportieren

    In Verbindung mit dem Mess- und Regelgerät DULCOMETER ® D1C, Chlor und den Dosierpumpen von ProMinent kann die Chlorkonzentration optimal geregelt werden. Typische Anwendung ist die Desinfektion von Trink-, Brauch-, Prozess- und Kühlwasser mit chlorhaltigen Desinfektionsmitteln insbesondere bei höheren pH-Werten bis 9,5.
  • Página 7: Aufbau Und Funktion

    Sensortyp nur wenig pH-abhängig. Das Primärsignal wird durch die Verstär- ker-Elektronik des Sensors in ein temperaturkorrigiertes Ausgangssignal ® 4-20 mA umgewandelt und im DULCOMETER D1C, Chlor zur Anzeige gebracht. Das Messsignal des CTE-Sensors folgt der DPD 4-Bestimmung (siehe Kap. 8.2). ® ProMinent...
  • Página 8 Aufbau und Funktion Abb. 1 Aufbau des Kabeldurchführung Sensors PG-7-Verschraubung Adapter Schutzkappe 5-Pol-Anschluss 2-Leiter-Anschluss Klemmscheibe Elektrodenschaft Schlauchdichtung Membrankappe Membran- schutzkappe Typ CTE 1-mA Typ CTE 1-DMT ® ProMinent...
  • Página 9 DPD-Bestimmung eingeschraubt werden (siehe Kalibrieren Kap. 8.2) Der Durchlaufgeber ist hydraulisch an den Messwasserstrom angeschlossen. Abb. 2 Messstelle ® DULCOMETER DULCOMETER STOP D1C Chlor Zweidraht- messleitung ® DULCOTEST CTE 1 Messwasser- Messwasser- eingang ausgang Durchlaufgeber DGMa Probe- nahmehahn ® ProMinent...
  • Página 10: Montieren

    Abb. 3) und den Elektrolyten lang- sam in einem Zug aus der Vorrats- Tülle flasche herausdrücken, dabei die Vorratsflasche stetig zurückziehen; die Kappe ist komplett gefüllt, wenn der Elektrolyt am unteren Gewindegang ansteht. Abb. 3: Membrankappe mit Elektrolytflasche ® ProMinent...
  • Página 11 O-Ring von unten über den Sensor bis zur Klemmscheibe schieben; eine Unterlegscheibe im DGM lassen. Sensor in DGM einführen und mit Klemmschraube fest anziehen, bis der O-Ring dichtet: Die richtige Einbautiefe des Sensors ist durch die Klemm- scheibe festgelegt. ® ProMinent...
  • Página 12: Installieren

    Zu geringe Versorgungsspannung verursacht einen fehlerhaften Messwert und kann in einem Regelkreis zu gefährlicher Überdosierung führen! Der Sensor CTE 1-mA ist ein Sensor mit passiver 4-20 mA-Zweileiter-Schnitt- stelle. Die Stromversorgung erfolgt von extern bzw. vom Mess- und Regel- ®...
  • Página 13 Installieren CTE 1-mA M12-Verschraubung lösen und das 2-adrige Kabel durchführen. Dabei die zweiadrige Messleitung in dem Sensor mit einer Überlänge von ca. 5 cm bevorraten. Die beiden Kabelenden abisolieren und wie aus Abb. 4 ersichtlich mit der Klemme verbinden (beiliegenden Schraubendreher verwenden). 1 = Plus, 2 = Minus (siehe Abb.
  • Página 14: In Betrieb Nehmen

    Einlaufzeit des Sensors abgewartet werden. Einlaufzeit Um einen stabilen Anzeigewert zu erreichen, benötigt der Sensor folgende Einlaufzeiten: Erstinbetriebnahme: 24 h nach Membranwechsel: 1-6 h Wiederinbetriebnahme: ca. 4-24 h Kalibrieren ACHTUNG Die in Abhängigkeit vom verwendeten Chlorungsmittel vor- geschriebene Kalibriermethode muss eingehalten werden. ® ProMinent...
  • Página 15 Falls noch nicht geschehen, Sensor in den Durchlaufgeber DLG III oder DGM einbauen (siehe Montieren, Kap. 6) Probenahme für die DPD-Messung durchführen. Diese muss in unmittel- barer Nähe zum Sensor erfolgen. Empfehlung: Benutzen Sie im Falle des Durchflussgebers DGM den Probenahmehahn (siehe Abb. 2 und Bestellhinweise, Kap.15) ® ProMinent...
  • Página 16 Kap. 8, Vollständiges Bedienmenü, Einstellmenü „Kalibrieren Chlor“). Überprüfen Sie nach einer Erstinbetriebnahme die Kalibrierung mittels DPD nach 24 Stunden. Folgende Kalibrierintervalle werden empfohlen • Trink-, Brauch-, Prozess-, Kühlwasser: abhängig von den speziellen Bedingungen (1-4 Wochen) • Schwimmbäder: wöchentlich • Whirlpools: täglich ® ProMinent...
  • Página 17: Sensor Warten

    Chlormessbesteck nach der DPD 4-Methode regelmäßig überprüfen. Wenn nötig, den Sensor neu kalibrieren (siehe Kalibrieren, Kap. 8.2). Ist die Kalibrierung nicht mehr möglich, muss die Membrankappe gewech- selt und die Kalibrierung wiederholt werden (siehe Kapitel 6, Montieren, 8.1 Einlaufzeit und 8.2 Kalibrieren). ® ProMinent...
  • Página 18: Fehler Beheben

    4) Kurzschluss in der Messleitung Kurzschluss aufspüren und beseitigen 5) Abstand zwischen Membran Membrankappe bis zum Anschlag und Elektrode zu groß zuschrauben 6) DPD-Chemikalien überaltert Neue DPD-Chemikalien ver- wenden, Kalibrieren wiederholen 7) pH-Wert < pH 6,5 pH-Wert anheben (pH 6,5-9,5) ® ProMinent...
  • Página 19 Durchfluss korrigieren (siehe Kap. 14, Technische Daten) 4) Luftblasen außen an der Membran Den Durchfluss innerhalb des erlaubten Bereichs erhöhen 5) Störende Substanzen Rücksprache mit ProMinent im Messwasser (Tenside, Öle, Alkohole, Korrosionsinhibitoren) 6) Messbereich wurde deutlich Membrankappe austauschen überschritten (siehe Kap. 6); Sensor einlaufen lassen (siehe Kap.
  • Página 20: Außer Betrieb Nehmen

    Falls aus dem Durchlaufgeber (DGM/DLG III) Flüssigkeit austritt, die bauseitig installierten Absperrhähne am Zu- und Ablauf schließen. • Vor dem Öffnen des DGM/DLG III die Sicherheitshinweise des Anlagenbetreibers beachten! • Beachten Sie zusätzlich alle Sicherheitshinweise in Kap. 6, Montieren! ® ProMinent...
  • Página 21: Reparieren

    Originalverpackung ein. Bereiten Sie den Sensor dafür vor (wie in Kap. 11, Außer Betrieb nehmen beschrieben). 13 Entsorgen ACHTUNG Beachten Sie die gültigen, nationalen Vorschriften. ProMinent Dosiertechnik GmbH, Heidelberg nimmt die dekontaminierten Altgeräte bei ausreichender Frankierung der Sendung zurück. ® ProMinent...
  • Página 22: Technische Daten

    Elektroden- schaft Versorgungs- spannung 16 - 24 V DC; min 35 mA bei 16 V DC Ausgangssignal CTE 1-mA: 4-20 mA, temperaturkorrigiert, unkalibriert und galvanisch nicht getrennt CTE 1-DMT: 0-1000 mV anschließend zum Anschluss an DULCOMETER ® Schutzart IP 65 Lagertemperatur zwischen 5 und 50 °C...
  • Página 23: Bestellhinweise

    (2 Membrankappen + 50 ml Elektrolyt) Bestell-Nr. 740048 • Zubehörset CTE 1 (0,5 ppm) (2 Membrankappen + 50 ml Elektrolyt) Bestell-Nr. 741277 • 1 Membrankappe kpl. (CTE 1-mA-2/5/10 ppm und CTE-DMT-10 ppm) Bestell-Nr. 792862 • 1 Membrankappe kpl. (CTE 1-mA-0,5 ppm) Bestell-Nr. 741274 Montageset •...
  • Página 24: Klemmenanschlussplan Dmt Chlor

    Klemmenanschlussplan DMT Chlor 17 Klemmenanschlussplan DMT Chlor Eingänge Ausgang Intern Extern ProMinent Universal- kabel Stromschleife 4-20 mA Speisespanung 12-40 V DC CTE 1-DMT-Sensor ACHTUNG • Beachten Sie die Anweisungen in der Betriebsanleitung des Vorort-Messumformer DMT Chlor! • Der Klemmenanschlussplan gilt nur für die DMT Variante! Er gibt an, wie man die Messleitung der DMT-Variante an den Vorort-Messumformer DMT Chlor anschließt.
  • Página 25 Installation ..............Operation ..............8.1 Run-in period ............8.2 Calibration ............. Maintenance ..............Troubleshooting ............Decommissioning ............Repairs ................Disposal ................ Technical Data ............... Ordering Information ............. Directives and standards complied with ....... Terminal connection diagram, DMT chlorine ....® ProMinent...
  • Página 26: How To Read The Operating Instructions

    CAUTION If safety guidelines are ignored, slight personal injury and damage to equipment may result! IMPORTANT If safety guidelines are ignored, damage to equipment may result! NOTE Working guidelines. ® ProMinent...
  • Página 27: Safety

    • Check that the delivery is complete by comparing your order with the delivery documents. Scope of delivery • 1 sensor CTE 1 • 1 bottle containing electrolyte (50 ml) and nozzle • 1 membrane cap for replacement • 1 operating instructions • 1 small screwdriver ® ProMinent...
  • Página 28: Storage And Transport

    ® the DULCOMETER D1C, chlorine measurement and control device and ProMinent metering pumps. Typical applications include the disinfection of potable, industrial, process and cooling water with chlorine-based disinfec- tants, particularly in the presence of higher pH values up to 9.5.
  • Página 29: Design And Function

    4-20 mA temperature-corrected output signal which is displayed in the ® DULCOMETER D1C, chlorine. The CTE sensor measurement signal follows DPD 4 detection (see section 8.2). ® ProMinent...
  • Página 30 Design and function Fig. 1 Sensor designs Cable bushing PG-7 threaded connector Adaptor Protective cap 5-pole connection 2-lead connection Tensioning ring Electrode shaft Rubber seal Membrane cap Protective membran cap Type CTE 1-mA Type CTE 1-DMT ® ProMinent...
  • Página 31 The in-line probe housing is hydraulically connected to the sample water flow. Fig. 2 Measurement station ® DULCOMETER DULCOMETER STOP D1C, chlorine Two-wire signal lead ® DULCOTEST CTE 1 Sample water Sample water inlet outlet DGMa in line probe housing Sampling tap ® ProMinent...
  • Página 32: Assembly

    (see figure 3). Slowly squeeze the Nozzle electrolyte out of the bottle in one steady stream, while continuously retracting the bottle. The cap is completely full when the electrolyte reaches the lower Fig. 3: Membrane cap with edge of the thread. electrolyte bottle ® ProMinent...
  • Página 33 Slide the O-ring from below over the sensor shaft up to the clamping ring. Leave a flat washer inside the DGM. Insert the sensor in the DGM and screw on the plug until the O-ring seals. The clamping ring determines the installation depth of the sensor. ® ProMinent...
  • Página 34: Installation

    The CTE 1-mA sensor has a passive 4-20 mA two-wire interface. The power supply is provided externally or by the controller. When the sensor is ®...
  • Página 35 Installation CTE 1-mA Loosen the M12 threaded connector and guide the 2-wire cable into the sensor. Leave an approx. 5 cm length of the 2-core measuring line in the sensor. Strip the two wires at their ends and connect them to the terminal block by using the screwdriver provided: 1 = plus, 2 = minus (see figure 4).
  • Página 36: Operation

    In order to achieve a stable display value the sensor should be run in for the following run in periods. First commissioning: 24 hour After changing membrane: 1-6 h Re-commissioning: approx. 4-24 hours Calibration IMPORTANT The calibration method specified depending on the used chlorination agent must be observed. ® ProMinent...
  • Página 37 Take a water sample for DPD-measurement. Sampling location has to be close to the installed sensor. Recommendation: use the sampling tap in the case of the DGM in-line probe housing (see Figure 2 and Ordering information, Chapter 15). ® ProMinent...
  • Página 38 After initial installation of the sensor, check the calibration by DPD- measurement 24 hours later. The following maintenance intervals based on experiences are recommended • Potable, industrial, process and cooling water depending on the specific conditions (1-4 weeks) • swimming pools: weekly • spa pools (whirlpools): daily ® ProMinent...
  • Página 39: Maintenance

    DPD-4 method using an appropriate chlorine measurement kit. If necessary, recalibrate the sensor (see Chapter 8.2 “Calibration”). If calibration cannot be carried out properly, the membrane cap must be replaced and calibration must be repeated (see Chapter 6 “Assembly”, 8.1 “Run-in time” and 8.2 “Calibration”). ® ProMinent...
  • Página 40: Troubleshooting

    Screw the membrane cap tightly electrode and membrane too great onto the electrode shaft, so that no gap can be seen anymore 6) DPD chemicals spent Use new DPD chemicals, repeat calibration 7) pH-value < pH 6.5 Raise pH-value (6.5-9.5) ® ProMinent...
  • Página 41 Examine sample water for is 3.0 - 4.0 mA consume chlorine interfering substances and remedy 3) Zero point has shifted Carry out zero point calibration (see Chapter 8.2) 4) Reference electrode defective* Return sensor to ProMinent for regeneration ® ProMinent...
  • Página 42 2) Reference electrode defective* Return sensor to ProMinent for regeneration * When the reference electrode turns shiny silver or white it must be regenerated. Brown-grey or yellow green coloration is normal.
  • Página 43: Decommissioning

    The sensor can be repaired only at the factory. Send it back in its original packaging. Prepare it for the repairs (as described in Chapter 11 “Decommissioning”). 13 Disposal IMPORTANT Please observe the applicable national regulations. You may return the decontaminated used equipment to ProMinent Dosiertechnik GmbH, Heidelberg, against prepaid postage. ® ProMinent...
  • Página 44: Technical Data

    Membrane cap: Electrode shaft: Supply voltage 16 - 24 V DC; min 35 mA at 16 V DC Output signal CTE 1-mA: 4-20 mA, temperature-corrected, un-calibrated and not electrically isolated CTE 1-DMT: 0-1000 mV after connection to DULCOMETER ® Enclosure rating...
  • Página 45: Ordering Information

    • Accessory kit CTE 1 (0.5 ppm) (2 membrane caps + 50 ml electrolyte) Order number 741277 • 1 membrane cap kit (CTE 1-mA-2/5/10 ppm and CTE-DMT-10 ppm) Order number 792862 • 1 membrane cap kit (CTE 1-mA-0.5 ppm) Order number...
  • Página 46: Terminal Connection Diagram, Dmt Chlorine

    Terminal connection diagram, DMT chlorine 17 Terminal connection diagram, DMT chlorine Input Output 4 3 2 1 Internal External ProMinent ® universal- cable Current loop 4-20 mA Supply voltage 16-40 V DC CTE 1-DMT sensor IMPORTANT • Follow the instructions given in the operating instruction manual for the DMT chlorine on-site measurement transducer.
  • Página 47 8.2 Etalonnage ............Maintenance de la sonde ..........Dépannage ..............Mise hors service ............Réparations ..............Elimination des déchets ..........Caractéristiques techniques ......... Références de commande ..........Directives et normes respectées ........Schéma des connexions de la DMT chlore ....® ProMinent...
  • Página 48: Remarques Sur La Lecture De Ce Mode D'emploi

    : PRUDENCE Le non-respect des consignes de sécurité peut entraîner des blessures corporelles et des dommages matériels légers. ATTENTION Le non-respect des consignes de sécurité peut créer des risques de dommages matériels. INFORMATION Consignes de travail. ® ProMinent...
  • Página 49: Sécurité

    • Vérifiez l’intégralité de la livraison sur la base de la commande et des documents de livraison. Etendue de la livraison • 1 sonde CTE 1 • 1 bouteille d’électrolyte (50 ml) et douille • 1 capuchon à membrane de rechange • 1 mode d’emploi • 1 petit tournevis ® ProMinent...
  • Página 50: Stockage Et Transport

    En liaison avec l’appareil de mesure et de régulation DULCOMETER D1C, chlore et les pompes doseuses de ProMinent, la concentration en chlore peut être dosée de façon optimale. Application typique : la désinfection de l’eau potable, des eaux usées et des eaux de process et de refroidissement contenant des désinfectants chloreux, surtout à...
  • Página 51: Structure Et Fonctionnement

    Le signal primaire est converti par l’électronique d’amplification de la sonde en un signal de sortie 4-20 mA à température corrigée puis indiqué sur ® le DULCOMETER D1C, chlore. Le signal de mesure de la sonde CTE est conforme à la directive DPD 4 (cf. chapitre 8.2). ® ProMinent...
  • Página 52 Passage de câble des sondes raccord vissé PG-7 Capuchon de Adaptateur protection Connexion Connexion 5 pôles 2 conducteurs Bague de serrage Tige d'électrode Joint Capuchon à membrane Capuchon de protection de la membrane Type CTE 1-mA Type CTE 1-DMT ® ProMinent...
  • Página 53 Fig. 2 Station de mesure ® DULCOMETER DULCOMETER STOP D1C, chlore Câble de mesure à deux fils ® DULCOTEST CTE 1 Entrée Sortie de l’eau de l’eau de mesure de mesure Chambre d’analyse DGMa Robinet de prise d’échantillons ® ProMinent...
  • Página 54: Montage

    à mesure du remplissage ; le capuchon est complètement rempli quand l’électrolyte atteint Fig. 3 : capuchon à membrane avec le pas inférieur du filetage. flacon d’électrolyte ® ProMinent...
  • Página 55 • Evitez les installations qui forment des bulles d’air dans l’eau de mesure ! Les bulles d’air adhérant au capuchon de membrane peuvent entraîner une valeur mesurée trop faible et ainsi un dosage incorrect dans un circuit de régulation. ® ProMinent...
  • Página 56: Montage / Installation

    éventuellement un surdosage dangereux dans un circuit de régulation ! La sonde CTE 1-mA est une sonde à interface passive à deux conducteurs 4-20 mA. L’alimentation de tension est assurée de manière externe par l’appareil de mesure et de régulation. Si vous raccordez le régulateur ®...
  • Página 57 Enlevez environ 5 cm de l’isolation extérieure du câble de sorte que les deux fils deviennent visibles. CTE 1-mA Dévissez le raccord M12 et faites passer le câble à 2 fils. Ce faisant, stockez un câble de mesure bifilaire dans la sonde présentant une surlongueur de 5 cm environ.
  • Página 58: Mise En Service

    : 24 h après un changement de membrane : 1 - 6 h nouvelle mise en service : env. 4 - 24 h Etalonnage ATTENTION Toujours respecter la méthode d’étalonnage prescrite en fonction de l’agent chlorant utilisé. ® ProMinent...
  • Página 59 à son mode d’emploi (voir le mode d’emploi du ® DULCOMETER D1C, chlore, chapitre 8, menu de commande complet, menu de réglage “Etalonnage point zéro chlore”). Remontez la sonde dans la chambre d’analyse (DGM, DLG III) comme décrit au chapitre 7. ® ProMinent...
  • Página 60 Nous recommandons les intervalles d’étalonnage suivants • Eau potable, eaux usées, eaux de process, eaux de refroidissement: en fonction des conditions spécifiques (1 à 4 semaines) • piscines : chaque semaine • bains à remous : chaque jour ® ProMinent...
  • Página 61: Maintenance De La Sonde

    DPD 4. Si nécessaire, recommencez l’étalonnage de la sonde (voir Etalonnage, paragraphe 8.2). Si l’étalonnage n’est plus possible, remplacez le capuchon à membrane et recommencez l’étalonnage (voir chap. 6, Montage, 8.1 Temps d’initialisa- tion et 8.2 Etalonnage). ® ProMinent...
  • Página 62: Dépannage

    14, dans l’eau, trouvez un remède caractéristiques techniques) 4) Court-circuit dans le câble Décelez le court-circuit et de mesure éliminez-le 5) Distance trop grande entre Vissez le capuchon à membrane la membrane et l’électrode jusqu’en butée ® ProMinent...
  • Página 63 4) Bulles d’air à l’extérieur de la Augmentez le débit dans la plage membrane admise 5) Substances perturbatrices Contactez ProMinent dans l’eau de mesure (agents tensioactifs, huiles, alcools, inhibiteurs de corrosion) 6) La plage de mesure est nettement Remplacez le capuchon à...
  • Página 64: Mise Hors Service

    • En cas d’urgence, débranchez d’abord le régulateur du secteur ! Si du liquide s’écoule de la chambre d’analyse (DGM/DLG III), fermez les robinets d’arrêt de l’alimentation et de l’évacuation installés par l’utilisateur. ® ProMinent...
  • Página 65: Réparations

    Préparez la sonde à cette intervention (comme indiqué au chapitre 11, Mise hors service). 13 Elimination des déchets ATTENTION Veuillez respecter les réglementations nationales en vigueur. Vous pouvez retourner les appareils usagés décontaminés, correctement affranchis, à ProMinent Dosiertechnik GmbH, Heidelberg. ® ProMinent...
  • Página 66: Caractéristiques Techniques

    16 - 24 V ; min. 35 mA à 16 V DC Signal de sortie CTE 1-mA: 4 - 20 mA, à température corrigée, non calibré et non séparé galvaniquement CTE 1-DMT: 0 - 1000 mV pour le branchement au DULCOMETER ®...
  • Página 67: Références De Commande

    (2 capuchons de membrane + 50 ml d’électrolyte) N° de commande 741277 • 1 capuchon de membrane complet (CTE 1-mA-2/5/10 ppm et CTE-DMT-10 ppm) N° de commande 792862 • 1 capuchon de membrane complet (CTE 1-mA-0,5 ppm) N° de commande...
  • Página 68: Directives Et Normes Respectées

    La fabrication est soumise à un haut degré de qualité qui est protégée par des directives et normes européennes. Une déclaration de conformité appropriée peut être demandée auprès de ProMinent. 17 Schéma des connexions de la DMT chlore Entrées Sortie Intérieur...
  • Página 69 Mantenimiento del sensor ..........Eliminación de fallos ............. Puesta fuera de servicio ..........Reparación ..............Eliminación de residuos ..........Datos técnicos .............. Instrucciones para el pedido ........Directivas y normas aplicadas ........Esquema de conexión de bornes DMT cloro ....® ProMinent...
  • Página 70: Indicaciones Para La Lectura De Las Instrucciones De Servicio

    CUIDADO En caso de incumplimiento de las indicaciones de seguridad existe peligro de daños personales y materiales ligeros. ATENCION En caso de incumplimiento de las indicaciones de seguridad existe peligro de daños materiales. OBSERVACION Instrucciones para el trabajo. ® ProMinent...
  • Página 71: Seguridad

    • Controle la integridad del envío según su pedido y los documentos del mismo. Volumen del suministro • 1 sensor CTE 1 • 1 botella de electrolito (50 ml) con boquilla • 1 cápsula de membrana de repuesto • 1 manual de instrucciones • 1 destornillador pequeño ® ProMinent...
  • Página 72: Almacenaje Y Transporte

    ® En combinación con el aparato de medición y regulación DULCOMETER D1C, cloro y las bombas dosificadoras ProMinent, se puede regular la concentración de cloro de forma óptima. Una aplicación típica es la desinfección de agua potable, aguas industriales, agua de procesos y agua de refrigeración con desinfectantes que contienen cloro, en especial con...
  • Página 73: Características Constructivas Y Función

    La señal primaria es transformada en una señal de salida con temperatura corregida por la electrónica de amplificación del sensor de 4-20 mA y se visualiza en el DULCOMETER ® D1C, cloro. La señal medida del sensor CTE sigue la determinación DPD 4 (ver cap. 8.2). ® ProMinent...
  • Página 74 Pasacables constructivas de Rosca PG-7 los sensores Caperuza de protección Adaptador Conexión de Conexión 5 polos 2 conductores Anillo fiador Portaelectrodos Junta de manguera Cápsula de membrana Caperuza protectora de la membrana Tipo CTE 1-DMT Tipo CTE 1-mA ® ProMinent...
  • Página 75 Fig. 2 Punto de medición DULCOMETER ® DULCOMETER STOP D1C, cloro Cable de medición bifilar DULCOTEST ® CTE 1 Entrada del Salida del agua de agua de medición medición Detector de paso DGMa Grifo de toma de muestras ® ProMinent...
  • Página 76: Montaje

    Fig. 3: Cápsula de la membrana con de la rosca. envase de electrolito ® ProMinent...
  • Página 77 Las burbujas de aire adheridas a la membrana del sensor pueden generar un valor de medición demasiado bajo y ser causa, por ello, de dosificaciones erróneas en un circuito de regulación. ® ProMinent...
  • Página 78: Montaje / Instalación

    El sensor CTE 1-mA es un sensor con interfaz de dos conductores pasivo 4-20 mA. La alimentación de corriente se realiza externa o por el aparato de ®...
  • Página 79 Quitar unos 5 cm del aislamiento exterior del cable de medición, de forma que sean visibles los dos conductores. CTE 1-mA Soltar la rosca M12 y hacer pasar el cable de 2 conductores. Dejar un exceso de cable de medición bifilar de aprox. 5 cm en el medidor (reserva).
  • Página 80: Puesta En Servicio

    Primera puesta en servicio: 24 h Después del cambio de la membrana: 1 - 6 h Nueva puesta en servicio: aprox. 4 - 24 h Calibración ATENCION Deberá observarse el método de calibración prescrito dependiente del agente clorante utilizado. ® ProMinent...
  • Página 81 ® servicio de DULCOMETER D1C, cloro, cap. 8, menú operativo completo, menú de ajuste “Calibración del punto cero cloro”). Montar el sensor nuevamente en el detector de paso (DGM, DLG III) de acuerdo con el cap. 7 Instalación. ® ProMinent...
  • Página 82 Controle, después de la primera puesta en servicio, la calibración con DPD después de 24 horas. Se recomiendan los intervalos de calibración siguientes: • Agua potable, industrial, de procesos, refrigeración: depende de las condiciones especiales (1-4 semanas) • Piscinas: semanal • Jacuzzis: diario ® ProMinent...
  • Página 83: Mantenimiento Del Sensor

    En caso necesario, calibrar nuevamente el sensor (véase calibración, cap. 8.2). Si ya no es posible la calibración, debe cambiarse la cápsula de la membrana y repetirse la calibración (véase capítulo 6 Montaje, 8.1 Tiempo de adaptación y 8.2 Calibración). ® ProMinent...
  • Página 84: Eliminación De Fallos

    Enroscar la cápsula de la mem- electrodo demasiado grande. brana hasta el tope. 6) Productos químicos de DPD Utilizar agentes químicos nuevos envejecidos. en el DPD, repetir la calibración. 7) Valor pH < pH 6.5 Elevar el valor pH (pH 6.5 - 9.5) ® ProMinent...
  • Página 85 4) Burbujas de aire en el exterior de Elevar el flujo dentro del volumen la membrana. permitido. 5) Sustancias perturbadoras en el Consultar con ProMinent. agua de medición (tensioactivos, aceites, alcoholes, inhibidores de corrosión). 6) Se ha sobrepasado claramente Cambiar la cápsula de la mem- el rango de medición...
  • Página 86 D1C. Para ello, lea el valor bajo “Punto cero” en el menú operativo completo, vea ® instrucciones de servicio de DULCOMETER D1C, cloro, cap. 8, en el menú de ajuste “Calibrar cloro”. Después no confirme con la tecla Intro, sino abandone el menú con al tecla “Volver”. ® ProMinent...
  • Página 87: Puesta Fuera De Servicio

    Lavar bien los electrodos con agua limpia caliente, de forma que no quede electrolito adherido, finalmente dejar secar. Para protección de los electrodos, enroscar sin apretar una cápsula de membrana nueva. Finalmente poner la caperuza de protección de la membrana. ® ProMinent...
  • Página 88: Reparación

    Prepare el sensor para el envío (como se describe en el cap. 11). 13 Eliminación de residuos ATENCION Observe por favor las reglamentaciones nacionales vigentes. ProMinent Dosiertechnik GmbH, Heidelberg acepta la devolución de los equipos viejos descontaminados si se envían con el franqueo suficiente. ® ProMinent...
  • Página 89: Datos Técnicos

    Tensión de alimentación 16 - 24 V CC; mín. 35 mA con 16 V CC Señal de salida CTE 1-mA: 4-20 mA, con temperatura corregida, sin calibrar y sin separación galvánica CTE 1-DMT: 0-1000 mV para conexión siguiente a DULCOMETER ®...
  • Página 90: Instrucciones Para El Pedido

    • Set de accesorios CTE 1 (0,5 ppm) (2 cápsulas de membrana + 50 ml electrolito) Referencia: 741277 • 1 cápsula de membrana completa (CTE 1-mA-2/5/10 ppm y CTE-DMT-10 ppm) Referencia: 792862 • 1 cápsula de membrana completa (CTE 1-mA-0,5 ppm)
  • Página 91: Directivas Y Normas Aplicadas

    La producción está sujeta a un alto standard de calidad, garantizado por normas y directivas europeas. La declaración de conformidad correspondiente se puede pedir a ProMinent. 17 Esquema de conexión de bornes DMT cloro Entradas...
  • Página 92 Anschriften- und Liefernachweis durch den Hersteller/ Addresses and delivery through manufacturer/ Adresses et liste des fournisseurs fournies par le constructeur/ Para informarse de las direcciones de los distribuidores, dirigirse al fabricante: ProMinent Dosiertechnik GmbH Im Schuhmachergewann 5-11 69123 Heidelberg · Germany Tel.: +49 6221 842-0 Fax: +49 6221 842-419 info@prominent.com...

Este manual también es adecuado para:

Cte 1-dmt

Tabla de contenido