Assembly - ProMinent DULCOTEST PER Serie Instrucciones De Servicio

Sensor de peróxido de hidrógeno
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Filling with electrolyte
ProMinent
®
6

Assembly

CAUTION
• Protective goggles and protective clothing should be worn
when handling water and solutions containing H
• Do not swallow the electrolyte. If electrolyte comes into
contact with the eyes or skin, rinse the affected areas
thoroughly with water! If reddening of the eyes occurs, con-
sult an eye specialist!
IMPORTANT
• Do not touch or damage the diaphragm and electrodes!
• Always keep electrolyte bottles tightly closed after use! Do
not transfer electrolyte into any other bottles or containers.
• The electrolyte should not be kept for more than 1 year! (See
label for use-by date).
• The diaphragm cap may only be used once!
Remove the diaphragm protection cap and unscrew the dia-
phragm cap from the electrode shaft.
Fill the diaphragm cap with electrolyte up to the lower thread, if
possible bubble-free.
If you wish to considerably reduce the running-in time, you must
expel the air between the gauze and the diaphragm (as the air is
expelled, the diaphragm is visible by reflecting through the
electrolyte as it is being filled).
There are two ways of doing this:
1.
Tap the diaphragm cap lightly from the side, from the bottom and
from the top with the sensor shaft until the air bubbles stop rising
(you can see this in good light).
2.
Fill the enclosed pipette with electrolyte from the diaphragm cap,
if possible up to the top. Please ensure that no air is taken in.
Take the opening of the pipette as close as possible to the
diaphragm (through the electrolyte) and place a couple of drops
on it from the pipette (do not release any air from the pipette
when doing this!)
IMPORTANT
After using the pipette, rinse thoroughly with water and store
in the original sensor packaging!
Assembly
O
!
2
2
31

Publicidad

Tabla de contenido
loading

Tabla de contenido