4 Acquiring CEPH Images
34
X-ray intensity (Hard, Normal, Soft) is according to the
operator's decision.
NOTE
Soft ≤ Normal ≤ Hard
If necessary, you may manually adjust these values using the arrows to the
right of each value.
When you click the Confirm button,
■ The Ready button will begin blinking to show that it is activate. This means
that the equipment is ready for X-Ray exposure.
9. Guide the patient to the equipment. Position the patient. Refer to section 4.2
Positioning the Patient.
6. Select X-ray intensity.
7. A default value for the tube voltage
(kVp) and current (mA) will be displayed
based on the gender and X-ray intensity.
8 . C l i c k C o n f i r m t o a c c e p t t h e
parameters.
■ The scan time and estimated DAP
(Dose Area Product) value are shown
on the main display for the intended
exposure.
PaX-i3D (PHT-6500) User Manual