4.2

Closed wireless network (group)

With a closed wireless network a main device monitors the
radio connections with all the group devices and all the group
devices monitor the radio connections with the main device.
When a device fails, e. g. as a result of a flat battery or removal
of a device, an alarm is triggered in the main device. If the main
device fails, an alarm is triggered in all the group devices.
NOTICE
i
i
For use of the following functions, the radio function
must be activated in the X-zone 5000.
4.2.1
Wireless network menu
NOTICE
i
i
Use the
and
OK
navigation in the menus.
To open the wireless network menu:
Press the
key.
OK
The
wireless
network
menu is displayed.
To navigate in the menus:
Open the wireless network menu /
key
OK
open the selected function
Select next function /
key
Show next page
Explanation of general symbols:
Main device
Group device
#x
Local station number of the X-zone
Group size
x
Action not possible
Confirmation indicator;
Press the
ALL
group or on the network.
Dräger X-zone 5000
keys on the X-am 5000/5600 for
Local
Net
Config
Test
key to confirm all alarms in the
OK
Explanation of wireless network menu symbols:
Show local measurements from the device
Local
(default display)
Show max. measurements of the group
Net
devices
Open group configuration
Config
Carrying out a connection test
Test
Explanation of group configuration symbols:
Provisionally deactivating group monitoring
Activating group monitoring
Forming a new group /
Adding a device to the group
Removing a device from the group
Terminating group
4.2.2
Displaying group devices
1. Press the
OK
2. Select
the group devices.
Legend:
1 Group device display
2 Local station number
3 Station number associated
with the measurement
display
4 Page display
5 Max. measurement
6 Gas name / unit
04733112.eps
NOTICE
i
i
For all gases, except for O
the group is displayed. The minimum value is
displayed for O
4.2.3
Forming a new group
A maximum of 15 group devices can be assigned to a main
device. Group affiliation is saved in the corresponding X-
zone 5000 until the group is terminated.
To form a new group:
1. Gather all the devices for the group in one area.
2. Switch on the devices (see "Switching on the instrument"
on page 38).
3. Select an X-zone 5000 as the main device.
4. Press the
OK
5. Select and confirm
6. Select and confirm
group devices.
Establishing a wireless connection
key on the main device to open the menu.
Net to display
1
2
#1
6
O
2
20.9
H
S
2
5
0
, the maximum value within
2
.
2
key on the main device to open the menu.
.
to start the search process for
Vol%
3
1
ppm
0
4
04433112.eps
43