11. Technical data
Model:
Veroval
Type:
GCE606
Measuring method:
oscillometric
Display range:
0 - 300 mmHg
Measuring range:
Systolic (SYS): 50–280 mmHg
Diastolic (DIA): 30–200 mmHg
Pulse: 40–199 pulse beats per minute
Displaying of correct values outside the
measuring range cannot be guaranteed.
Display unit:
1 mmHg
Technical
Cuff pressure: ± 3 mmHg
measuring accuracy:
Pulse: ± 5% of the displayed pulse rate
Clinical measuring
complies with the requirements of DIN
accuracy:
EN 1060-4;
Korotkoff validation method: Phase I (SYS),
Phase V (DIA)
Operating mode:
continuous mode
Nominal voltage:
DC 6V
BPU 22
®
Power supply:
Expected service life:
Battery capacity:
Protection against
electric shock:
Protection against
harmful ingress
of water or solid
materials:
Inflation pressure:
Automatic switch-off
function:
Cuff:
Memory capacity:
58
58
4× 1.5 V alkaline manganese (AA/LR06)
batteries
20,000 measurements
about 1,000 measurements
Medical-electrical equipment with internal
power supply (only when batteries are
used); applied part: type BF
IP21 (not protected against moisture)
about 190 mmHg for first measurement
1 minute after measurement / otherwise
30 sec.
Veroval
cuff for BPU 22,
®
cuff for arm circumference of 22 to 42 cm
2 × 100 measurements with average of all
measurements and morning / evening aver-
age of the last 7 days