• Keep away from direct sunlight and heat. Store the test strip bottle
in a dry, cool place.
• Record the "date opened" on the bottle label. Discard the bottle and
any remaining test strip after 6 months from date of opening.
• Make sure you are testing in an environment between 50°F~104°F
and relative humidity less than 85%. Your monitor will not work
outside these ranges.
Step 3:
Insert a test strip to turn on the monitor. If the talking feature is on the
monitor will say "Please apply blood onto the test strip." A flashing
blood drop and arrow will appear in the monitor display screen.
According to the level of control solution press and release the bottom
button for (LOW-L1). Press and release the top button for (HIGH-L2).
L1/L2 will appear on the display screen.
The ctl icon will appear next to the test strip icon.
If the talking feature is on the monitor will say "L1/L2
control solution test please apply Control Solution
onto the test strip".
Step 4:
Gently guide the tip of the strip into the drop of the
control solution. When the strip and monitor have
received the control solution the monitor will begin
to play a melody. The monitor will start to count
down.
26
Blood Glucose Monitoring System