Control Solution Trouble Shooting
If your control solution testing is out of range (too high or too low), it
may be caused by the following:
Possible Causes
• Wrong brand of Control
Solution being used
• Control Solution not at room
temperature
• Expired or contaminated
control solution
• Damaged test strip
• Monitor malfunction
28
Blood Glucose Monitoring System
What you can do
• Make sure you are using Embrace
TALK Control Solution.
• Make sure the testing environment
is between 50°F to 104°F, <85%
RH.
• Check the expiration and open
bottle date on both the control
solution and test strips.
(Test strips should be used within 6
months from date opened. Control
solution should be used within 3
months from date opened).
Repeat testing using a new test
strip. If the result is still out of
range, use a new bottle of control
solution and retest.
• If problem persists, call Omnis
Health@1-877-979-5454.