What
What the
you see
Monitor says
"Error 1
Monitor
Failure"
"Error 3 used
test strip"
"Error 4 not
enough
sample"
"Temperature
is out of
range"
Blood Glucose Monitoring System
What it
means
The monitor
electronics are
damaged
Used or damaged
test strip
Not enough
sample in the
test strip for the
monitor to read.
Temperature out
of range
What you should do
Remove and put back in
the batteries and turn the
monitor back on.
If problems persists, call
us at 1-877-979-5454
Repeat the test with a
new test strip. Remember
to add sample only when
you see the prompt on
the display screen or the
monitor verbally prompts
you to add sample.
Remove the test strip and
repeat the test with a new
test strip. See "Testing
Your Blood Glucose" on
page 29.
Move the Monitor into
an area that is between
50°F~104°F and relative
humidity less than 85%,,
and allow 10 to 15
minutes for the monitor's
temperature to get back
with in range.
45