Descargar Imprimir esta página

Dräger Interlock XT Instrucciones De Uso página 29

Ocultar thumbs Ver también para Interlock XT:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 48
Display
Violation Reset
Handset error / Contact Service !
Controlbox error / Contact Service !
Check cable and reconnect !
Controlbox mismatch
SW-version "OK"
(No display texts on handset, but signal
lamps functioning)
(No display messages on handset)
Dräger Interlock XT
Cause
Too many violations have been detected. Call an authorized Dräger Interlock Service Center in
Fault of handset.
Fault of control box.
No connection between handset and
control box.
Controlbox and handset have different
software versions.
Temperature of the handset too low.
No connection between handset and
control box.
Remedy
due time.
Switch ignition off/on and repeat the test.
If message reappears: call an authorized Dräger
Interlock Service Center.
Switch ignition off/on and repeat the test.
If message reappears: call an authorized Dräger
Interlock Service Center.
Check plugs of the connecting cable and connect again
with handset and control box.
If message reappears: call an authorized Dräger
Interlock Service Center.
After acknowledging with button "OK" the software
version may be seen (e.g. Rev. 1.9).
Call an authorized Dräger Interlock Service Center.
Perform test as usual guided by the signal lamps.
Detach handset and take it to a place with a moderate
temperature. Connect handset again and make
measurement.
Check plugs of the connecting cable, disconnect plug
from handset and reconnect again.
If again no messages are shown: call an authorized
Dräger Interlock Service Center.
Faults, cause and remedy