Technical data
12
Technical data
Measuring principle
Electrochemical DrägerSensor
Measuring range
up to 2.38 mg/L
Ambient conditions
for operation 
–45 °C to 85 °C (–49 °F to 185 °F)
and storage
20 % to 98 % r.h.
600 to 1100 hPa
No influence of changes in altitude on
the measurement result
Sensitivity drift
typically 1 % of measured value / month
Display
Graphic display
Calibration interval
typically 12 months
Dimensions (H x W x D)
Handset
approx. 150 mm x 70 mm x 40 mm
Control box
approx. 115 mm x 105 mm x 40 mm
Weight
Handset
approx. 175 g
Control box
approx. 320 g
Voltage supply
12 V to 24V
Current consumption
maximum
<2 A
stand-by
<20 mA
30
Relays for the lead starter re-
<16 A, continuous
lays
<40 A, peak
<48 V
Lead ignition
5 V to 48 V
Waiting time to ready for
measurement
< 10 seconds
≥20 °C (68 °F)
< 3 minutes
-40 °C (-40 °F)
Approvals
Radio interference
(Directive 2006/28/EC)
EN 50436-1: 2005
EN 50436-2: 2007
Dräger Interlock XT