Adjustment – Easy Calibration
8.1
Adjustment
The nominal volume or volume to be tested is set,
standard mode Pipetting (PIP), e.g. 200.0 µl (for
procedure, see PIP mode). Ex: Volume according to
volume test 301.3 µl.
a.
Bring up CAL mode: Press and hold the MENU key
1x
(> 3 sec) until CAL appears.
>3s
The display reads ‚off'.
'CAL' blinks.
b.
Activate CAL mode: Press one of the arrow buttons
to activate the CAL mode.
The display changes from ‚off' to ‚on'.
1x
c. Confirm CAL mode: Press the Enter key.
The display now shows the set pipetting volume.
'CAL' blinks.
d.
Setting the volume: Use the arrow keys (+/-) to set
the volume, which was previously determined and
tested.
'CAL' blinks.
1x
e. Confirm volume: Press the Enter key.
The display shows the tested and corrected
volume.
The CAL symbol is continuously displayed to
confirm that an adjustment has been made.
ErgoOne
'CAL' continues to blink.
E Multi-Channel Pipettes · User Manual · 23
®