Disassembly of ErgoOne
The three main components of the pipetting unit can be easily separated and disassembled for
maintenance, cleaning or replacing parts. The procedure is explained in detail on the following
pages.
Changing the O-rings on the individual shafts is described in detail in the instructions included
with the replacement part.
f.
Mount piston: Insert cleaned or new piston.
Tighten the piston nut again using the installation
tool.
g.
Assemble the piston and shaft unit: Loosen the
shafts by half a turn. Then place the shaft unit on
the piston unit and fasten it. Then screw on the
shafts.
h.
Assemble pipetting unit: Reassemble the
pipetting unit and mount it on the handle. Check
the instrument for leaks, free movement, and
precision.
NOTICE!
The pipetting unit must be screwed onto the handle in a
counter-clockwise direction until it clicks into place. Do
not pull the pipetting unit downwards.
i.
Carry out reference run: Carry out reference run
(rEF).
E Multi-Channel pipettes 1250 µl
®
ErgoOne
E Multi-Channel Pipettes · User Manual · 29
®