Idiomas disponibles

Conducted RF
3 V for
IEC 61000-4-6
0.15–80 MHz;
6 V in ISM
and am-
ateur fre-
quencies
between
0.15 and
80 MHz
Radiated RF
385 kHz,
27 V/m
IEC 61000-4-3
450 MHz,
28 V/m
710 MHz,
745 MHz,
780 MHz
9 V/m
810 MHz,
870 MHz,
930 MHz
28 V/m
1720 MHz,
1845 MHz,
1970 MHz
28 V/m
2450 MHz,
28 V/m
5240 MHz,
5500 MHz,
5785 MHz
9 V/m
Information about electromagnetic immunity
3 V for
0.15–80 MHz;
6 V in ISM
and am-
ateur fre-
quencies
between
0.15 and
80 MHz
385 kHz,
27 V/m
450 MHz,
28 V/m
710 MHz,
745 MHz,
780 MHz
9 V/m
810 MHz,
870 MHz,
930 MHz
28V/m
1720 MHz,
1845 MHz,
1970 MHz
28 V/m
2450 MHz,
28 V/m
5240 MHz,
5500 MHz,
5785 MHz
9 V/m
Portable and mobile RF
communications equip-
ment should be used
no closer to any part of
the ABO 523 model, in-
cluding cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended separation
3.5
distance
���� = [
] √ ����
3.5
����
���� = [
] √ ����
1
����
7
1
���� = [
] √ ����
����
7
80 MHz to 800 MHz
���� = [
] √ ����
1
����
1
800 MHz to 2.7 GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, a should be less than
the compliance level in each
frequency range.
The product may experi-
ence interference when
it is near devices with the
following symbol: