Descargar Imprimir esta página

Pari eBase Controller Instrucciones De Uso página 22

Ocultar thumbs Ver también para eBase Controller:

Publicidad

Idiomas disponibles

Idiomas disponibles

Disinfection
WARNING:
Disinfect the nebuliser handset parts and
the aerosol head at least once a day, e.g.,
immediately after cleaning at the end of
the day (a nebuliser handset cannot be
disinfected effectively unless it has been
cleaned first).
NOTE:
Do not use a microwave to disinfect the
aerosol head. The possibility of product
components being damaged cannot be
ruled out.
Using a standard commercial
disinfector for baby bottles (so-called
steam sterilizer)
• To ensure effective disinfection, use a
thermal disinfector with an operating
time of at least 6 minutes. With regard to
performing the disinfection, the duration
of the disinfection process and the
quantity of water required, follow the
instructions for use of the disinfector you
are using.
18
CAUTION:
Inadequate disinfection promotes the
growth of bacteria and thus increases
the risk of infection. If disinfection is not
carried out for the time specified for the
device concerned, it will not be effective.
Therefore, do not switch the device off
before the disinfection cycle is complete.
Also, make sure that the device is clean
and check it regularly for proper operation.
• Take the parts out of the disinfector
immediately after disinfection is com-
pleted. Place them on a dry, clean sur-
face and allow them to dry completely.
CAUTION:
Check the nebuliser handset parts after
every disinfection. Replace all broken,
deformed and heavily discoloured parts.
eBase
®
Controller - 2022-09

Publicidad

loading

Este manual también es adecuado para:

678g1020