6.4.3 (METHOD D) Use of gluteraldehyde type
disinfectant: immerse the medicine tray (2) and
the accessories (3, 4, 5) in a solution in accord-
ance with the product manufacturer's instruc-
tions (e.g. Korsolex basic, 4% for 15 minutes).
6.4.3.a Leave the accessories immersed for the
period of time indicated on the disinfectant
packaging, and associated with the concentra-
tion chosen for the preparation of the solution.
6.4.3.b Take the accessories and rinse them
5
with plenty of warm drinking water.
2
6.4.3.c DRYING: carefully perform all the op-
4
erations described in section 6.1.2. of the user
3
manual.
6.4.3.d Dispose of the solution according to the
disinfectant manufacturer's instructions.
7 - REASSEMBLING THE APPLIANCE
7.1 Insert the medicine tray and mesh (2) in the
appliance case (1) using the appropriate guide
(A).
Ensure the accessory shank is on the same side
as the button.
A
7.2 Push until the two parts are fully coupled.
35